Philips Healthcare SureSigns VS2+ Model 863278 SureSigns VS2+ NBP and SpO2 Model 863279 SureSigns VS2+ NBP, SpO2 and Temperature Product Usage: The SureSigns VS2+ vital signs monitor is for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility.
Published February 20, 2013 · Recall initiated: January 14, 2013
Summary
FDA records medical device Class II recall Z-0824-2013, published February 20, 2013 and initiated January 14, 2013. Recalling firm: Philips Healthcare Inc.; Status: Terminated; Reported quantity: 3248 units; Lot: Serial Numbers: CN14800428, CN14800429, CN14800430, CN14800431, CN14800432, CN14800433, CN14800434, CN14800435, CN14800436, CN14800437, CN14800438, CN14800439, CN14800440, CN14800756, CN14800763, CN14800764, CN14800765, CN14800772, CN14800773, CN14800822, CN14800823, CN14800824, CN14800979, CN14801332, CN21201423, CN21201625, CN21202029, CN21202030, CN21202031, CN21202032, CN21202033, CN21202034, CN21202234, CN21202239, CN21202240, CN21202241, CN21202242, CN21202243, CN21202317, CN21202320, CN21202347, CN21202348, CN21202349, CN21202350, CN21202351, CN21202376, CN21202435, CN21202606, CN21202607, CN21202862, CN21202863, CN21202883, CN21203006, CN21203007, CN21203008, CN21203009, CN21203010, CN21203073, CN21203180, CN21203304, CN21203307, CN21203308, CN21203309, CN21203310, CN21203311, CN21203312, CN21203707, CN21203781, CN21203794, CN21203795, CN21203796, CN21203797, CN21203949, CN21203950, CN21203951, CN21203952, CN21203953, CN21204204, CN21204205, CN21204206, CN21204207, CN21204208, CN21204209, CN21204210, CN21204211, CN21203057, CN14800484, CN14800744, CN14800745, CN14800746, CN21203044, CN21203045, CN21203068, CN21203069, CN21203070, CN21203071, CN14800485, CN14800486, CN14800749, CN14800750, CN14800751, CN14800752, CN14800770, CN14800771, CN14800893, CN21201756, CN21201757, CN21201758, CN21201759, CN21201760, CN21202164, CN21202165, CN21202166, CN21202167, CN21202168, CN21202169, CN21202170, CN21202171, CN21202172, CN21202173, CN21202174, CN21202175, CN21202176, CN21202177, CN21202178, CN21202202, CN21202610, CN21202864, CN21203209, CN21203210, CN21203639, CN14801200, CN14801201, CN14801202, CN14801203, CN14801204, CN14801205, CN14801206, CN14801207, CN14801208, CN14801209, CN14801210, CN14801212, CN14801213, CN14801217, CN14801278, CN14801279, CN14801280, CN14801281, CN21201576, CN21201613, CN21202313, CN21202367, CN21202368, CN21202371, CN21202657, CN21203072, CN21203416, CN21203887, CN21203961, CN21203970, CN21204148, CN21204318, CN21202393, CN21202394, CN21203746, CN21203747, CN21203748, CN21203749, CN21201924, CN21201925, CN21201926, CN21201927, CN21201928, CN21201929, CN21201930, CN21201931, CN21201932, CN21201942, CN21202263, CN21202264, CN21202669, CN21202727, CN21202728, CN21202729, CN21202730, CN21202819, CN21202842, CN21202843, CN21202844, CN21202845, CN21202846, CN21202847, CN21202848, CN21202849, CN21202850, CN21202851, CN21202852, CN21202853, CN21202854, CN21202855, CN21202856, CN21202926, CN21202927, CN21202928, CN21202929, CN21202930, CN21202931, CN21202932, CN21202933, CN21202934, CN21202935, CN21202936, CN21202937, CN21202938, CN21202939, CN21202940, CN21202941, CN21202942, CN21202943, CN21202944, CN21202945, CN21202946, CN21202947, CN21202948, CN21202949, CN21202950, CN21202951, CN21202952, CN21202953, CN21202954, CN21202955, CN21202956, CN21202957, CN21202958, CN21202959, CN21202960, CN21202961, CN21202962, CN21202963, CN21202964, CN21202965, CN21202966, CN21202967, CN21202968, CN21202969, CN21202970, CN21202971, CN21202972, CN21202973, CN21202974, CN21202975, CN21202976, CN21202977, CN21202978, CN21202979, CN21202980, CN21202981, CN21202982, CN21202983, CN21202984, CN21202985, CN21202986, CN21202987, CN21202988, CN21202989, CN21202990, CN21202991, CN21203197, CN21203198, CN21203199, CN21203200, CN21203201, CN21203202, CN21203203, CN21203204, CN21203205, CN21203206, CN21203340, CN21203341, CN21203342, CN21203343, CN21203344, CN21203345, CN21203346, CN21203347, CN21203348, CN21203349, CN21203350, CN21203351, CN21203352, CN21203353, CN21203354, CN21203355, CN21203356, CN21203357, CN21203358, CN21203359, CN21203360, CN21203725, CN21203726, CN21203727, CN21203728, CN21203729, CN21203730, CN21203731, CN21203740, CN21204314, CN21202373, CN14800478, CN14800481, CN14800637, CN14800699, CN14800700, CN14800891, CN14800892, CN14800968, CN14800969, CN14800976, CN14800977, CN14800978, CN14800985, CN21201578, CN21201579, CN21201749, CN21201750, CN21201751, CN21201752, CN21202193, CN21202214, CN21202215, CN21202216,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
VS2+ screen may lock up and stop monitoring during patient use
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Healthcare Inc.
- Distributor
- Worldwide Distribution - US Nationwide and the countries of: AUSTRALIA, AUSTRIA, BELGIUM,CANADA, CHILE, DENMARK, FRANCE GERMANY, ICELAND, INDIA, IRELAND,ISRAEL,ITALY, JORDAN, KENYA KUWAIT, LEBANON, MALAYSIA, NAMIBIA, NETHERLANDS, NEW ZEALAND NORWAY,PALESTINE, POLAND,PORTUGAL, SAUDI ARABIA, SINGAPORE SLOVENIA, SOUTH AFRICA, SOUTH KOREA,SPAIN,SWEDEN, SWITZERLAND THAILAND, TRINIDAD AND TOBAGO, UNITED ARAB EMIRATES UNITED KINGDOM
- Lot
- Serial Numbers: CN14800428, CN14800429, CN14800430, CN14800431, CN14800432, CN14800433, CN14800434, CN14800435, CN14800436, CN14800437, CN14800438, CN14800439, CN14800440, CN14800756, CN14800763, CN14800764, CN14800765, CN14800772, CN14800773, CN14800822, CN14800823, CN14800824, CN14800979, CN14801332, CN21201423, CN21201625, CN21202029, CN21202030, CN21202031, CN21202032, CN21202033, CN21202034, CN21202234, CN21202239, CN21202240, CN21202241, CN21202242, CN21202243, CN21202317, CN21202320, CN21202347, CN21202348, CN21202349, CN21202350, CN21202351, CN21202376, CN21202435, CN21202606, CN21202607, CN21202862, CN21202863, CN21202883, CN21203006, CN21203007, CN21203008, CN21203009, CN21203010, CN21203073, CN21203180, CN21203304, CN21203307, CN21203308, CN21203309, CN21203310, CN21203311, CN21203312, CN21203707, CN21203781, CN21203794, CN21203795, CN21203796, CN21203797, CN21203949, CN21203950, CN21203951, CN21203952, CN21203953, CN21204204, CN21204205, CN21204206, CN21204207, CN21204208, CN21204209, CN21204210, CN21204211, CN21203057, CN14800484, CN14800744, CN14800745, CN14800746, CN21203044, CN21203045, CN21203068, CN21203069, CN21203070, CN21203071, CN14800485, CN14800486, CN14800749, CN14800750, CN14800751, CN14800752, CN14800770, CN14800771, CN14800893, CN21201756, CN21201757, CN21201758, CN21201759, CN21201760, CN21202164, CN21202165, CN21202166, CN21202167, CN21202168, CN21202169, CN21202170, CN21202171, CN21202172, CN21202173, CN21202174, CN21202175, CN21202176, CN21202177, CN21202178, CN21202202, CN21202610, CN21202864, CN21203209, CN21203210, CN21203639, CN14801200, CN14801201, CN14801202, CN14801203, CN14801204, CN14801205, CN14801206, CN14801207, CN14801208, CN14801209, CN14801210, CN14801212, CN14801213, CN14801217, CN14801278, CN14801279, CN14801280, CN14801281, CN21201576, CN21201613, CN21202313, CN21202367, CN21202368, CN21202371, CN21202657, CN21203072, CN21203416, CN21203887, CN21203961, CN21203970, CN21204148, CN21204318, CN21202393, CN21202394, CN21203746, CN21203747, CN21203748, CN21203749, CN21201924, CN21201925, CN21201926, CN21201927, CN21201928, CN21201929, CN21201930, CN21201931, CN21201932, CN21201942, CN21202263, CN21202264, CN21202669, CN21202727, CN21202728, CN21202729, CN21202730, CN21202819, CN21202842, CN21202843, CN21202844, CN21202845, CN21202846, CN21202847, CN21202848, CN21202849, CN21202850, CN21202851, CN21202852, CN21202853, CN21202854, CN21202855, CN21202856, CN21202926, CN21202927, CN21202928, CN21202929, CN21202930, CN21202931, CN21202932, CN21202933, CN21202934, CN21202935, CN21202936, CN21202937, CN21202938, CN21202939, CN21202940, CN21202941, CN21202942, CN21202943, CN21202944, CN21202945, CN21202946, CN21202947, CN21202948, CN21202949, CN21202950, CN21202951, CN21202952, CN21202953, CN21202954, CN21202955, CN21202956, CN21202957, CN21202958, CN21202959, CN21202960, CN21202961, CN21202962, CN21202963, CN21202964, CN21202965, CN21202966, CN21202967, CN21202968, CN21202969, CN21202970, CN21202971, CN21202972, CN21202973, CN21202974, CN21202975, CN21202976, CN21202977, CN21202978, CN21202979, CN21202980, CN21202981, CN21202982, CN21202983, CN21202984, CN21202985, CN21202986, CN21202987, CN21202988, CN21202989, CN21202990, CN21202991, CN21203197, CN21203198, CN21203199, CN21203200, CN21203201, CN21203202, CN21203203, CN21203204, CN21203205, CN21203206, CN21203340, CN21203341, CN21203342, CN21203343, CN21203344, CN21203345, CN21203346, CN21203347, CN21203348, CN21203349, CN21203350, CN21203351, CN21203352, CN21203353, CN21203354, CN21203355, CN21203356, CN21203357, CN21203358, CN21203359, CN21203360, CN21203725, CN21203726, CN21203727, CN21203728, CN21203729, CN21203730, CN21203731, CN21203740, CN21204314, CN21202373, CN14800478, CN14800481, CN14800637, CN14800699, CN14800700, CN14800891, CN14800892, CN14800968, CN14800969, CN14800976, CN14800977, CN14800978, CN14800985, CN21201578, CN21201579, CN21201749, CN21201750, CN21201751, CN21201752, CN21202193, CN21202214, CN21202215, CN21202216,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMobile Diagnostic X-Ray System with Digital PanelApril 12, 2017
- FDAPhilips IntellVue Info Center iX 866023; PIIC iX Upgrade 866024; IntellVue Server iX Upgrade 866026; PIICJune 18, 2014
- FDAPhilips Easy Upgrade DR, Code No: 712086 This system is used for making X-Ray exposures for diagnostics.September 4, 2013
- FDAPhilips Healthcare MobileDiagnostwDR system. Mobile Digital radiography X-Ray System Product Usage:August 28, 2013
- FDASLIMDIA Revolution capsules, 30-count capsules per bottle, Yerba Naturals, 19685 Varda Ave., Buena Park, CA.February 27, 2013
- FDAGE Healthcare, CARESCAPE Monitor B650. The CARESCAPE Monitor B650 is a multi-parameter patient monitorFebruary 20, 2013
Questions about this recall
Was Philips Healthcare SureSigns VS2+ Model 863278 SureSigns VS2+ NBP and SpO2 Model 863279 SureSigns VS2+ NBP, SpO2 and Temperature Product Usage: The SureSigns VS2+ vital signs monitor is for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility. recalled?
Yes. FDA published recall Z-0824-2013 on February 20, 2013. Reported hazard: VS2+ screen may lock up and stop monitoring during patient use
What should I do if I have this product?
Voluntary: Firm initiated · Letter