ELECTROLASE(R) HYFRECATOR(R) ELECTRODES - BLUNT, REF/Catalog Number 7-101-8BX, Disposable Hyfrecator(R) Tips, Sterile, Rx Only --- CONMED CORPORATION USA. Intended to be used as active RF electrodes for desiccation, fulguration and coagulation.

Published February 27, 2013 · Recall initiated: January 28, 2013

Summary

FDA records medical device Class II recall Z-0827-2013, published February 27, 2013 and initiated January 28, 2013. Recalling firm: ConMed Corporation; Status: Terminated; Reported quantity: Domestic Distribution: 97,450 units; Foreign Distribution: 6,800 units; Lot: Beginning Lot Code 120628X, Ending Lot Code 120917X. [Lot codes on boxes and packaging contain a lot code in the following format: 120917X where 12 represents the year (2012), 09 represents the month (September), 17 represents the day of the month, and X represents the manufacturing shift code (1, 2, 3, 4,5,6 or 9)]. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Some devices had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
ConMed Corporation
Distributor
Worldwide Distribution-USA (nationwide) and the countries of China, Dubai and Indonesia.
Lot
Beginning Lot Code 120628X, Ending Lot Code 120917X. [Lot codes on boxes and packaging contain a lot code in the following format: 120917X where 12 represents the year (2012), 09 represents the month (September), 17 represents the day of the month, and X represents the manufacturing shift code (1, 2, 3, 4,5,6 or 9)].
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ELECTROLASE(R) HYFRECATOR(R) ELECTRODES - BLUNT, REF/Catalog Number 7-101-8BX, Disposable Hyfrecator(R) Tips, Sterile, Rx Only --- CONMED CORPORATION USA. Intended to be used as active RF electrodes for desiccation, fulguration and coagulation. recalled?

Yes. FDA published recall Z-0827-2013 on February 27, 2013. Reported hazard: Some devices had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products.

What should I do if I have this product?

Voluntary: Firm initiated · Letter