Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number AGNE20J; b) ADULT CRANIOTOMY PACK, kit number LLCR99; c) LAMINECTOMY PACK, kit number LLLC26; d) KIT, CRANIOTOMY, kit number MMCR11J; e) KIT, NEURO SPINE, kit number MMNS20H; f) IR NEURO PACK, kit number OSIN77; g) NEURO SAH, kit number SANE32H; h) NEURO SAH, kit number SANE32I; i) S78FCH CRANI PACK (PS 132432), kit number UICR10U; j) CRANI PACK (PS039498), kit number UICR33AW; k) Laminectomy, kit number UILM77AR; l) DIAGNOSTIC NEURO IR TRAY, kit number WENE21D; m) DIAGNOSTIC NEURO IR TRAY, kit number WENE21F; n) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H; o) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H-01; p) NEURO IR TRAY, kit number WENI56F; q) NEURO IR TRAY, kit number WENI56H; r) NEURO IR TRAY, kit number WENI56H-01
Published February 7, 2024 · Recall initiated: November 29, 2023
Summary
FDA records medical device Class II recall Z-0827-2024, published February 7, 2024 and initiated November 29, 2023. Recalling firm: American Contract Systems, Inc.; Status: Ongoing; Reported quantity: 2717 units; Lot: a) AGNE20J, UDI/DI 00191072183187, Lot Numbers: 934231, exp. 10/30/2023; 900231, exp. 4/10/2024; 990231, exp. 1/10/2024; 687221, exp. 11/9/2023; 720221, exp. 10/7/2023; 757221, exp. 8/31/2023; 870221, exp. 5/10/2023; 702221, exp. 10/25/2023; 795221, exp. 7/24/2023; b) LLCR99, UDI/DI 00191072162755, Lot Numbers: 756221, exp. 9/1/2023; 720221, exp. 10/7/2023; 654221, exp. 12/12/2023; 949231, exp. 2/20/2024; c) LLLC26, UDI/DI 00191072171900, Lot Numbers: 749221, exp. 9/8/2023; 695221, exp. 11/1/2023; 997231, exp. 1/3/2024; 920231, exp. 3/21/2024; d) MMCR11J, UDI/DI none, Lot Numbers: 744221, exp. 9/13/2023; 939231, exp. 3/2/2024; 876231, exp. 5/4/2024; e) MMNS20H, UDI/DI none, Lot Numbers: 963231, exp. 2/6/2024; 959231, exp. 2/10/2024; f) OSIN77, UDI/DI 00191072170705, Lot Numbers: 780221, exp. 8/8/2023; 728221, exp. 9/29/2023; 668221, exp. 11/28/2023; 991231, exp. 1/9/2024, 960231, exp. 2/9/2024; 939231, exp. 3/2/2024; UDI/DI 00191072190031, Lot Numbers: 912231, exp. 3/29/2024; 903231, exp. 4/7/2024; g) SANE32H, UDI/DI none, Lot Numbers: 838221, exp. 6/11/2023; 889221, exp. 4/21/2023; h) SANE32I, UDI/DI none, Lot Numbers: 673221, exp. 11/23/2023; 730221, exp. 9/27/2023; 784221, exp. 8/4/2023; 911231, exp. 3/30/2024; 969231, exp. 1/31/2024; i) UICR10U, UDI/DI none, Lot Numbers: 990231, exp. 1/10/2024; 921231, exp. 3/20/2024; 885231, exp. 4/25/2024; j) UICR33AW, UDI/DI none, Lot Numbers: 658221, exp. 12/8/2023; 931231, exp. 3/10/2024; 876231, exp. 5/4/2024; k) UILM77AR, UDI/DI 00191072139184, Lot Numbers: 953231, exp. 2/16/2024; 905231, exp. 4/5/2024; l) WENE21D, UDI/DI 00191072167958, Lot Numbers: 839221, exp. 6/10/2023; m) WENE21F, UDI/DI 00191072167958, Lot Numbers: 819221, exp. 6/30/2023; n) WENE21H, UDI/DI 00191072179326, Lot Numbers: 729221, exp. 9/28/2023; 686221, exp. 11/10/2023; o) WENE21H-01, UDI/DI 00191072179326, Lot Numbers: 980231, exp. 1/20/2024; p) WENI56F, UDI/DI 00191072169341, Lot Numbers: 799221, exp. 7/20/2023; q) WENI56H, UDI/DI 00191072179333, Lot Numbers: 700221, exp. 10/27/2023; r) WENI56H-01, UDI/DI 00191072179333, Lot Numbers: 652221, exp. 12/14/2023; 996231, exp. 1/4/2024. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- American Contract Systems, Inc.
- Distributor
- US Nationwide distribution.
- Lot
- a) AGNE20J, UDI/DI 00191072183187, Lot Numbers: 934231, exp. 10/30/2023; 900231, exp. 4/10/2024; 990231, exp. 1/10/2024; 687221, exp. 11/9/2023; 720221, exp. 10/7/2023; 757221, exp. 8/31/2023; 870221, exp. 5/10/2023; 702221, exp. 10/25/2023; 795221, exp. 7/24/2023; b) LLCR99, UDI/DI 00191072162755, Lot Numbers: 756221, exp. 9/1/2023; 720221, exp. 10/7/2023; 654221, exp. 12/12/2023; 949231, exp. 2/20/2024; c) LLLC26, UDI/DI 00191072171900, Lot Numbers: 749221, exp. 9/8/2023; 695221, exp. 11/1/2023; 997231, exp. 1/3/2024; 920231, exp. 3/21/2024; d) MMCR11J, UDI/DI none, Lot Numbers: 744221, exp. 9/13/2023; 939231, exp. 3/2/2024; 876231, exp. 5/4/2024; e) MMNS20H, UDI/DI none, Lot Numbers: 963231, exp. 2/6/2024; 959231, exp. 2/10/2024; f) OSIN77, UDI/DI 00191072170705, Lot Numbers: 780221, exp. 8/8/2023; 728221, exp. 9/29/2023; 668221, exp. 11/28/2023; 991231, exp. 1/9/2024, 960231, exp. 2/9/2024; 939231, exp. 3/2/2024; UDI/DI 00191072190031, Lot Numbers: 912231, exp. 3/29/2024; 903231, exp. 4/7/2024; g) SANE32H, UDI/DI none, Lot Numbers: 838221, exp. 6/11/2023; 889221, exp. 4/21/2023; h) SANE32I, UDI/DI none, Lot Numbers: 673221, exp. 11/23/2023; 730221, exp. 9/27/2023; 784221, exp. 8/4/2023; 911231, exp. 3/30/2024; 969231, exp. 1/31/2024; i) UICR10U, UDI/DI none, Lot Numbers: 990231, exp. 1/10/2024; 921231, exp. 3/20/2024; 885231, exp. 4/25/2024; j) UICR33AW, UDI/DI none, Lot Numbers: 658221, exp. 12/8/2023; 931231, exp. 3/10/2024; 876231, exp. 5/4/2024; k) UILM77AR, UDI/DI 00191072139184, Lot Numbers: 953231, exp. 2/16/2024; 905231, exp. 4/5/2024; l) WENE21D, UDI/DI 00191072167958, Lot Numbers: 839221, exp. 6/10/2023; m) WENE21F, UDI/DI 00191072167958, Lot Numbers: 819221, exp. 6/30/2023; n) WENE21H, UDI/DI 00191072179326, Lot Numbers: 729221, exp. 9/28/2023; 686221, exp. 11/10/2023; o) WENE21H-01, UDI/DI 00191072179326, Lot Numbers: 980231, exp. 1/20/2024; p) WENI56F, UDI/DI 00191072169341, Lot Numbers: 799221, exp. 7/20/2023; q) WENI56H, UDI/DI 00191072179333, Lot Numbers: 700221, exp. 10/27/2023; r) WENI56H-01, UDI/DI 00191072179333, Lot Numbers: 652221, exp. 12/14/2023; 996231, exp. 1/4/2024
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: Banner University Med Ctr. Product Name: Cardiac Cath Lab Pack-170278 Model/Catalog Number:April 2, 2025
- FDAHand Pack, AKHD97C; Medical convenience kitOctober 2, 2024
- FDAVaginal Delivery, ASDV25B; Medical convenience kitOctober 2, 2024
- FDAIR Procedure Pack, CEAT19W; Medical convenience kitOctober 2, 2024
- FDACustom procedural convenience kits and trays, general hospital use, labeled as: a) Baby PICC Kit, kit numberFebruary 7, 2024
- FDACustom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LOWER PELVICFebruary 7, 2024
Questions about this recall
Was Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number AGNE20J; b) ADULT CRANIOTOMY PACK, kit number LLCR99; c) LAMINECTOMY PACK, kit number LLLC26; d) KIT, CRANIOTOMY, kit number MMCR11J; e) KIT, NEURO SPINE, kit number MMNS20H; f) IR NEURO PACK, kit number OSIN77; g) NEURO SAH, kit number SANE32H; h) NEURO SAH, kit number SANE32I; i) S78FCH CRANI PACK (PS 132432), kit number UICR10U; j) CRANI PACK (PS039498), kit number UICR33AW; k) Laminectomy, kit number UILM77AR; l) DIAGNOSTIC NEURO IR TRAY, kit number WENE21D; m) DIAGNOSTIC NEURO IR TRAY, kit number WENE21F; n) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H; o) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H-01; p) NEURO IR TRAY, kit number WENI56F; q) NEURO IR TRAY, kit number WENI56H; r) NEURO IR TRAY, kit number WENI56H-01 recalled?
Yes. FDA published recall Z-0827-2024 on February 7, 2024. Reported hazard: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.