Back to search
FDAFDA · Food and drugs·Recall number Z-0829-2024

Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LABOR & DELIVERY PACK, kit number HILD52F; b) LABOR & DELIVERY PACK, kit number HILD52G

Published February 7, 2024 · Recall initiated: November 29, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
American Contract Systems, Inc.
Distributor
US Nationwide distribution.
Lot
a) HILD52F, UDI/DI none, Lot Numbers: 846221, exp. 6/3/2023; 807221, exp. 7/12/2023; 959231, exp. 11/30/2023; b) HILD52G, UDI/DI none, Lot Numbers: 885231, exp. 4/25/2024
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card