adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.

Published December 31, 2014 · Recall initiated: November 7, 2014

Summary

FDA records medical device Class II recall Z-0836-2015, published December 31, 2014 and initiated November 7, 2014. Recalling firm: Ion Beam Applications S.A.; Status: Terminated; Reported quantity: 1; Lot: PAT.109 (US), Treatment room 4. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ion Beam Applications S.A.
Distributor
US Distribution in PA only.
Lot
PAT.109 (US), Treatment room 4
Risk classification
Class II
Area (FDA)
device
Country
Belgium

Related recalls

Questions about this recall

Was adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device. recalled?

Yes. FDA published recall Z-0836-2015 on December 31, 2014. Reported hazard: Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du

What should I do if I have this product?

Voluntary: Firm initiated · Letter