adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.

Published December 31, 2014 · Recall initiated: November 7, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Ion Beam Applications S.A.
Distributor
US Distribution in PA only.
Lot
PAT.109 (US), Treatment room 4
Risk classification
Class II
Area (FDA)
device
Country
Belgium

Sell products like this?