Back to search
FDAFDA · Food and drugs·Recall number Z-0839-2014

DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Foot/Ankle Blanket NS, REF T656NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.

Published February 5, 2014 · Recall initiated: September 17, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
DeRoyal Industries Inc
Distributor
Nationwide Distribution.
Lot
Lot Numbers: 32901129, 32980912, 33015541, 33024746, 33105239
Risk classification
Class II
Area (FDA)
device
Country
United States