Back to search
FDAFDA · Food and drugs·Recall number Z-0845-2019

Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer

Published February 27, 2019 · Recall initiated: January 17, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer. A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Distributor
US and Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia, Syrian Arab Republic
Lot
UPN: 00643169121973 Lot/Serial Numbers: JVD000376P JVD000415P JVD000242P JVD000262P JVD000264P JVD000272P JVD000278P JVD000283P JVD000289P JVD000291P JVD000293P JVD000295P JVD000300P JVD000311P JVD000316P JVD000320P JVD000330P JVD000332P JVD000333P JVD000334P JVD000337P JVD000338P JVD000344P JVD000346P JVD000348P JVD000350P JVD000351P JVD000352P JVD000354P JVD000356P JVD000357P JVD000359P JVD000360P JVD000361P JVD000362P JVD000363P JVD000364P JVD000367P JVD000369P JVD000373P JVD000377P JVD000379P JVD000381P JVD000382P JVD000383P JVD000384P JVD000388P JVD000390P JVD000392P JVD000394P JVD000407P JVD000408P JVD000409P JVD000410P JVD000411P JVD000412P JVD000413P JVD000417P JVD000419P JVD000420P JVD000421P UPN: 0643169166974 Lot/Serial Numbers: JVD000331P JVD000426P JVD000430P UPN: 00643169166981 Lot/Serial Numbers: JVD000273P JVD000275P JVD000299P JVD000307P JVD000339P UPN: 00643169166998 Lot/Serial Numbers: JVD000245P JVD000265P JVD000297P JVD000301P JVD000313P JVD000387P JVD000391P JVD001090P JVD001137P UPN: 00643169167001 Lot/Serial Numbers: JVD000353P JVD000395P UPN: 00643169167018 Lot/Serial Numbers: JVD000386P JVD000389P JVD000425P JVD000432P JVD000269P JVD000290P JVD000326P UPN: 00643169167025 Lot/Serial Numbers: JVD000439P JVD000308P JVD000385P JVD001089P JVD000370P UPN: 00643169167049 Lot/Serial Numbers: JVD000322P JVD000378P JVD000418P JVD000434P UPN: 00643169167063 Lot/Serial Numbers: JVD000298P JVD000305P JVD000318P JVD000342P JVD000355P UPN: 00643169167070 Lot/Serial Numbers: JVD000374P JVD000292P JVD000324P JVD000343P JVD000371P JVD000247P JVD000257P JVD000261P JVD000287P JVD000372P JVD000399P JVD000405P JVD000414P UPN: 00643169167087 Lot/Serial Numbers: JVD000253P JVD000263P JVD000284P JVD000304P JVD000317P JVD000319P JVD000393P JVD000398P JVD000402P JVD001031P JVD001091P UPN: 00643169167094 Lot/Serial Numbers: JVD000282P JVD000271P JVD000397P JVD000404P UPN: 00643169167100 Lot/Serial Numbers: JVD000267P JVD001004P JVD000294P JVD000314P UPN: 00643169225732 Lot/Serial Numbers: JVD001067P JVD001068P JVD001098P JVD001112P JVD001146P JVD001147P JVD001008P JVD001016P JVD001063P JVD001064P JVD001065P JVD000525P JVD001002P JVD000474P JVD000489P JVD000501P JVD000523P JVD000463P JVD000473P JVD000484P JVD000459P UPN: 00643169225749 Lot/Serial Numbers: JVD000251P JVD000266P JVD000285P UPN: 00643169225756 Lot/Serial Numbers: JVD000347P JVD000288P UPN: 00643169225763 Lot/Serial Numbers: JVD000259P JVD000260P UPN: 00643169225770 Lot/Serial Numbers: JVD000244P JVD000254P JVD000256P JVD000258P JVD000276P JVD000309P JVD000310P UPN: 00643169369030 Lot/Serial Numbers: JVD001032P JVD001271P JVD000279P JVD000437P JVD001027P JVD002984P JVD002985P JVD002986P JVD002988P JVD002989P JVD002991P JVD002992P JVD002996P JVD002997P JVD002999P JVD003001P JVD003002P JVD003004P JVD003006P JVD001389P JVD001465P JVD000255P JVD001024P JVD001186P JVD001212P JVD001226P JVD001235P JVD001242P JVD001292P JVD001406P JVD001454P JVD001481P JVD001502P JVD001345P JVD001185P JVD001267P JVD001347P JVD000401P JVD001026P JVD001110P JVD001152P JVD001203P JVD001247P JVD001311P JVD001507P JVD000321P JVD000336P JVD000424P JVD000428P JVD000461P JVD000487P JVD000493P JVD000528P JVD001000P JVD001102P JVD001104P JVD001106P JVD001115P JVD001138P JVD001157P JVD001171P JVD001178P JVD001179P JVD001187P JVD001191P JVD001202P JVD001204P JVD001208P JVD001214P JVD001221P JVD001246P JVD001260P JVD001298P JVD001299P JVD001306P JVD001308P JVD001314P JVD001349P JVD001387P JVD001419P JVD001431P JVD001434P JVD001447P JVD001617P JVD001018P JVD001448P JVD001289P JVD001415P JVD001023P JVD001253P JVD001466P JVD001467P JVD001283P JVD001512P JVD001057P JVD00143
Risk classification
Class II
Area (FDA)
device
Country
United States