FDAFDA · Food and drugs·Recall number Z-0849-2015

Illumina MiSeqDx Universal Kit 1.0, PN 15039608 The Illumina MiSeqDx Universal Kit 1.0 is a set of reagents and consumables used in the processing of human genomic DNA samples derived from peripheral whole blood, and in the subsequent targeted sequencing of the resulting sample libraries. User-supplied analyte specific reagents are required for the preparation of libraries targeting specific genomic regions of interest. The MiSeqDx Universal Kit 1.0 is intended for use with the MiSeqDx instrument.

Published December 31, 2014 · Recall initiated: November 13, 2014

Summary

FDA records medical device Class II recall Z-0849-2015, published December 31, 2014 and initiated November 13, 2014. Recalling firm: Illumina Inc; Status: Terminated; Reported quantity: 13 units; Lot: Serial No. M70130, M70189, M70108, M70165, M70166, M70175, M70172, M70103. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It was discovered a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or deletions at specific genomic locations with respect to the primer.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Illumina Inc
Distributor
US Distribution to the states of AZ, GA, CA, OH, and VA., and one Internationally to France.
Lot
Serial No. M70130, M70189, M70108, M70165, M70166, M70175, M70172, M70103.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Illumina MiSeqDx Universal Kit 1.0, PN 15039608 The Illumina MiSeqDx Universal Kit 1.0 is a set of reagents and consumables used in the processing of human genomic DNA samples derived from peripheral whole blood, and in the subsequent targeted sequencing of the resulting sample libraries. User-supplied analyte specific reagents are required for the preparation of libraries targeting specific genomic regions of interest. The MiSeqDx Universal Kit 1.0 is intended for use with the MiSeqDx instrument. recalled?

Yes. FDA published recall Z-0849-2015 on December 31, 2014. Reported hazard: It was discovered a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or deletions at specific genomic locations with respect to the primer.

What should I do if I have this product?

Voluntary: Firm initiated · Letter