Hanger-Bar 2P 45cm
Published April 6, 2022 · Recall initiated: November 1, 2021
Summary
FDA records medical device Class II recall Z-0852-2022, published April 6, 2022 and initiated November 1, 2021. Recalling firm: Human Care USA, Inc.; Status: Ongoing; Reported quantity: 15 devices; Lot: Part Number 72755 UPC Code: 881608800403 EAN Code: 7331599999714 Lot Number: 33164504. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Human Care USA, Inc.
- Distributor
- U.S. Nationwide distribution in the states of CA, IL, LA, MA, NV, OH, OK, PA, TX, UT, VA, and WV.
- Lot
- Part Number 72755 UPC Code: 881608800403 EAN Code: 7331599999714 Lot Number: 33164504
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATETRACAINE 1% Tetracaine HCI Injection, USP, 20mg/2mL (10mg/mL), 10 x 2ml Single Use Vials per box, Rx only,April 13, 2022
- FDAPAPAVERINE HYDROCHLORIDE Injection, USP, 60 mg/2mL (30 mg/mL), packaged as a) 25 x 2mL Single Use Vials perApril 13, 2022
- FDAPHENOBARBITAL Sodium Injection, USP, 65mg/mL, packaged as a) 25 x 1 mL Vials per box (NDC 42494-415-25) and b)April 13, 2022
- FDABALLARD* Multi-Access Port Catheter, Double Swivel Y Elbow, Adult, 14 French; Sterile; Single UseApril 6, 2022
- FDARevogene, For In Vitro Diagnostic use in performing nucleic acid testing in clinical laboratories. MeridianApril 6, 2022
- FDACARA (Computer Assisted Retinal Analysis), Product ID: +B543CARA0/$$+7V4.7.2/16D202111179April 6, 2022
Questions about this recall
Was Hanger-Bar 2P 45cm recalled?
Yes. FDA published recall Z-0852-2022 on April 6, 2022. Reported hazard: The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.
What should I do if I have this product?
Voluntary: Firm initiated · Letter