LASEREDGE Knives, CRESCENT BLADE ANGLED TIP BEVEL UP (6/BOX), REF/PRODUCT CODE E7510AT, STERILE, Rx Only Product Usage: The LaserEdge Knives are intended for ophthalmic surgery. There is no therapeutic use and they are available to surgeons for use in procedures.
Published March 14, 2018 · Recall initiated: March 3, 2017
Summary
FDA records medical device Class II recall Z-0853-2018, published March 14, 2018 and initiated March 3, 2017. Recalling firm: Bausch & Lomb Inc Irb; Status: Terminated; Reported quantity: 1,236,749 knives in total; Lot: none. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The complaints database was reviewed and there have been an atypical volume of dull blades complaints for various lots. out of an abundance of precaution, Bausch + Lomb is voluntarily recalling all product codes SKU manufacturing without the electro-polish control and penetration force testing.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Bausch & Lomb Inc Irb
- Distributor
- US Nationwide Distribution
- Lot
- none
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDALASEREDGE Knives, 1.1MM ARROW KNIFE ANGLED (6/BOX), REF/PRODUCT CODE E7592, STERILE, Rx Only Product Usage:March 14, 2018
- FDALASEREDGE Knives, CLEAR 3.0 MM CLEAR CORNEAL ANGLED KNIFE (6/BOX), REF/PRODUCT CODE E7529, STERILE, Rx OnlyMarch 14, 2018
- FDALASEREDGE Knives, 3.2MM SLIT KNIFE ANGLED (6/BOX), REF/PRODUCT CODE E7557A, STERILE, Rx Only Product Usage:March 14, 2018
- FDALASEREDGE Knives, 2.4 MM SLIT KNIFE ANGLED DOUBLE BEVEL, REF/PRODUCT CODE E7549ADB, STERILE, Rx Only ProductMarch 14, 2018
- FDALASEREDGE Knives, CRESCENT BLADE BEVELED UP ANGLED (6/BOX), REF/PRODUCT CODE E7510, STERILE, Rx Only ProductMarch 14, 2018
- FDALASEREDGE Knives, 20 GAUGE MVR BLADE (6/BOX), REF/PRODUCT CODE E7520, STERILE, Rx Only Product Usage: TheMarch 14, 2018
Questions about this recall
Was LASEREDGE Knives, CRESCENT BLADE ANGLED TIP BEVEL UP (6/BOX), REF/PRODUCT CODE E7510AT, STERILE, Rx Only Product Usage: The LaserEdge Knives are intended for ophthalmic surgery. There is no therapeutic use and they are available to surgeons for use in procedures. recalled?
Yes. FDA published recall Z-0853-2018 on March 14, 2018. Reported hazard: The complaints database was reviewed and there have been an atypical volume of dull blades complaints for various lots. out of an abundance of precaution, Bausch + Lomb is voluntarily recalling all product codes SKU manufacturing without the electro-polish control and penetration force testing.
What should I do if I have this product?
Voluntary: Firm initiated · Letter