Bausch & Lomb Stellaris PC w/ Laser Upgrade BL1433 (Stellaris PC w v3 Software) with BL2272 (Laser Upgrade kit) Stellaris PC Vision Enhancement System with integrated laser module. System is used for the surgical treatment/removal of cataractous lens, vitreous removal and certain treatments of the posterior portion of the eye.

Published December 31, 2014 · Recall initiated: November 4, 2014

Summary

FDA records medical device Class II recall Z-0863-2015, published December 31, 2014 and initiated November 4, 2014. Recalling firm: Bausch & Lomb Inc; Status: Terminated; Reported quantity: 79; Lot: Serial numbers: SPC00146, SPC00148, SPC00181, SPC00362, SPC00384, SPC00393, SPC00403, SPC00406, SPC00418, SPC00422, SPC00424, SPC00427, SPC00466, SPC00469, SPC00500, SPC00504, SPC00524, SPC00525, SPC00566, SPC00636, SPC00738, SPC00771, SPC00785, SPC00793, SPC00802, SPC00839, SPC00851, SPC00889, SPC00926, SPC00966, SPC00969, SPC01002, SPC01035, SPC01159, SPC01167, SPC01213, SPC01218, SPC01224, SPC01263, SPC01267, SPC01300, SPC01347, SPC01405, SPC01442, SPC01446, SPC01447, SPC01449, SPC01466, SPC01474, SPC01492, SPC01542, SPC01573, SPC01608, SPC01629, SPC01653, SPC01655, SPC01666, SPC01667, SPC01678, SPC01681, SPC01687, SPC01696, SPC01699, SPC01707, SPC01732, SPC01742, SPC01746, SPC01749, SPC01750, SPC01752, SPC01754, SPC01756, SPC01760, SPC01816, SPC01817, SPC01819, SPC01820, SPC01843 and SPC01884. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Bausch & Lomb Inc
Distributor
Worldwide Distribution - USA Distribution including to the states of AL, CA, CT, FL, GA, HI, IL, IN, KS, MI, MN, MO, MS, NH, NY, OH, OK, PA, TX, WI and WV., and to the countries of Austria, Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Ireland, Italy, Jordan, Malaysia, Morocco, Netherlands, Norway, Portugal, Singapore, Spain, Sweden, Switzerland, Turkey and the United Arab Emirates.
Lot
Serial numbers: SPC00146, SPC00148, SPC00181, SPC00362, SPC00384, SPC00393, SPC00403, SPC00406, SPC00418, SPC00422, SPC00424, SPC00427, SPC00466, SPC00469, SPC00500, SPC00504, SPC00524, SPC00525, SPC00566, SPC00636, SPC00738, SPC00771, SPC00785, SPC00793, SPC00802, SPC00839, SPC00851, SPC00889, SPC00926, SPC00966, SPC00969, SPC01002, SPC01035, SPC01159, SPC01167, SPC01213, SPC01218, SPC01224, SPC01263, SPC01267, SPC01300, SPC01347, SPC01405, SPC01442, SPC01446, SPC01447, SPC01449, SPC01466, SPC01474, SPC01492, SPC01542, SPC01573, SPC01608, SPC01629, SPC01653, SPC01655, SPC01666, SPC01667, SPC01678, SPC01681, SPC01687, SPC01696, SPC01699, SPC01707, SPC01732, SPC01742, SPC01746, SPC01749, SPC01750, SPC01752, SPC01754, SPC01756, SPC01760, SPC01816, SPC01817, SPC01819, SPC01820, SPC01843 and SPC01884
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Bausch & Lomb Stellaris PC w/ Laser Upgrade BL1433 (Stellaris PC w v3 Software) with BL2272 (Laser Upgrade kit) Stellaris PC Vision Enhancement System with integrated laser module. System is used for the surgical treatment/removal of cataractous lens, vitreous removal and certain treatments of the posterior portion of the eye. recalled?

Yes. FDA published recall Z-0863-2015 on December 31, 2014. Reported hazard: A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.

What should I do if I have this product?

Voluntary: Firm initiated · Telephone