smiths medical Medfusion Model 4000 syringe pump
Published February 7, 2024 · Recall initiated: December 19, 2023
Summary
FDA records medical device Class I recall Z-0864-2024, published February 7, 2024 and initiated December 19, 2023. Recalling firm: Smiths Medical ASD Inc.; Status: Ongoing; Reported quantity: 60146 pumps; Lot: All Medfusion Model 4000 with software versions: v1.0.0, v1.1.0, v1.1.1, v1.1.2, v1.5.0, v1.5.1, 1.6.0, v1.6.1, v1.6.4, v2.3, v2.4, v2.5, and all versions prior to v1.6.5. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Medfusion Model 4000 syringe infusion pumps, with the following software versions, may have the following issues: 1. Delivery During Motor Not Running High Priority Alarm (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 2. Re-administered Loading Dose (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 3. Incorrect Critical Data Failure Alarm, 4. Interruption of Bolus or Loading Dose Delivery (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 5. Incorrect Total Bolus/Loading Dose Displayed (v1.1.0, v1.1.1, v1.1.2), 6. Volume Limit Before Bolus/Loading Dose Complete (v1.1.2), 7. Drug Library Lower Limit Displayed Incorrectly (v1.6.0, v1.6.1), 8. Depleted Battery Alarm (All versions previous to v1.6.5), 9. Loss of Wireless Connectivity (v1.5.0, v1.5.1, v1.6.0, v1.6.1, v1.6.4), 10. PharmGuard Server Password (v2.3, v2.4, v2.5). Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software installed on your pumps.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Smiths Medical ASD Inc.
- Distributor
- US nationwide, Canada, Great Britain, Bermuda, Australia, Germany, Singapore, Saudi Arabia, Lebanon, and Malaysia
- Lot
- All Medfusion Model 4000 with software versions: v1.0.0, v1.1.0, v1.1.1, v1.1.2, v1.5.0, v1.5.1, 1.6.0, v1.6.1, v1.6.4, v2.3, v2.4, v2.5, and all versions prior to v1.6.5.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of theNovember 6, 2024
- FDAENT RES, 100 ML, Non-Sterile W CLAMP & FEMALE LUER, Product Code 68-2231-28October 30, 2024
- FDACADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24JPOctober 30, 2024
- FDACADD Medication Cassette Reservoir, 100mL, Product Code 21-7002-24October 30, 2024
- FDAsmiths medical Medfusion Model 3500 Syringe pumpFebruary 7, 2024
- FDAGE Voluson IC9-RS intracavitary probes, Model Number J48691PJ, Diagnostic ultrasound imaging and fluid flowFebruary 7, 2024
Questions about this recall
Was smiths medical Medfusion Model 4000 syringe pump recalled?
Yes. FDA published recall Z-0864-2024 on February 7, 2024. Reported hazard: Medfusion Model 4000 syringe infusion pumps, with the following software versions, may have the following issues: 1. Delivery During Motor Not Running High Priority Alarm (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 2. Re-administered Loading Dose (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 3. Incorrect Critical Data Failure Alarm, 4. Interruption of Bolus or Loading Dose Delivery (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 5. Incorrect Total Bolus/Loading Dose Displayed (v1.1.0, v1.1.1, v1.1.2), 6. Volume Limit Before Bolus/Loading Dose Complete (v1.1.2), 7. Drug Library Lower Limit Displayed Incorrectly (v1.6.0, v1.6.1), 8. Depleted Battery Alarm (All versions previous to v1.6.5), 9. Loss of Wireless Connectivity (v1.5.0, v1.5.1, v1.6.0, v1.6.1, v1.6.4), 10. PharmGuard Server Password (v2.3, v2.4, v2.5). Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software installed on your pumps.
What should I do if I have this product?
Voluntary: Firm initiated · Letter