smiths medical Medfusion Model 4000 syringe pump

Published February 7, 2024 · Recall initiated: December 19, 2023

Summary

FDA records medical device Class I recall Z-0864-2024, published February 7, 2024 and initiated December 19, 2023. Recalling firm: Smiths Medical ASD Inc.; Status: Ongoing; Reported quantity: 60146 pumps; Lot: All Medfusion Model 4000 with software versions: v1.0.0, v1.1.0, v1.1.1, v1.1.2, v1.5.0, v1.5.1, 1.6.0, v1.6.1, v1.6.4, v2.3, v2.4, v2.5, and all versions prior to v1.6.5. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medfusion Model 4000 syringe infusion pumps, with the following software versions, may have the following issues: 1. Delivery During Motor Not Running High Priority Alarm (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 2. Re-administered Loading Dose (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 3. Incorrect Critical Data Failure Alarm, 4. Interruption of Bolus or Loading Dose Delivery (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 5. Incorrect Total Bolus/Loading Dose Displayed (v1.1.0, v1.1.1, v1.1.2), 6. Volume Limit Before Bolus/Loading Dose Complete (v1.1.2), 7. Drug Library Lower Limit Displayed Incorrectly (v1.6.0, v1.6.1), 8. Depleted Battery Alarm (All versions previous to v1.6.5), 9. Loss of Wireless Connectivity (v1.5.0, v1.5.1, v1.6.0, v1.6.1, v1.6.4), 10. PharmGuard Server Password (v2.3, v2.4, v2.5). Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software installed on your pumps.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smiths Medical ASD Inc.
Distributor
US nationwide, Canada, Great Britain, Bermuda, Australia, Germany, Singapore, Saudi Arabia, Lebanon, and Malaysia
Lot
All Medfusion Model 4000 with software versions: v1.0.0, v1.1.0, v1.1.1, v1.1.2, v1.5.0, v1.5.1, 1.6.0, v1.6.1, v1.6.4, v2.3, v2.4, v2.5, and all versions prior to v1.6.5.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was smiths medical Medfusion Model 4000 syringe pump recalled?

Yes. FDA published recall Z-0864-2024 on February 7, 2024. Reported hazard: Medfusion Model 4000 syringe infusion pumps, with the following software versions, may have the following issues: 1. Delivery During Motor Not Running High Priority Alarm (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 2. Re-administered Loading Dose (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 3. Incorrect Critical Data Failure Alarm, 4. Interruption of Bolus or Loading Dose Delivery (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 5. Incorrect Total Bolus/Loading Dose Displayed (v1.1.0, v1.1.1, v1.1.2), 6. Volume Limit Before Bolus/Loading Dose Complete (v1.1.2), 7. Drug Library Lower Limit Displayed Incorrectly (v1.6.0, v1.6.1), 8. Depleted Battery Alarm (All versions previous to v1.6.5), 9. Loss of Wireless Connectivity (v1.5.0, v1.5.1, v1.6.0, v1.6.1, v1.6.4), 10. PharmGuard Server Password (v2.3, v2.4, v2.5). Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software installed on your pumps.

What should I do if I have this product?

Voluntary: Firm initiated · Letter