Varian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Published February 27, 2013 · Recall initiated: January 28, 2013
Summary
FDA records medical device Class II recall Z-0867-2013, published February 27, 2013 and initiated January 28, 2013. Recalling firm: Varian Medical Systems, Inc. Oncology Systems; Status: Terminated; Reported quantity: 1168 units; Lot: H480044 , H481353 , H482733 , H484585 , H485948 , H480045 , H481354 , H482734 , H484634 , H486024 , H480105 , H481433 , H482765 , H484673 , H486032 , H480106 , H481436 , H482827 , H484677 , H486093 , H480159 , H481437 , H482858 , H484680 , H486094 , H480160 , H481451 , H482859 , H484745 , H486095 , H480247 , H481452 , H482860 , H484746 , H486114 , H480301 , H481453 , H482888 , H484747 , H486140 , H480373 , H481454 , H482889 , H484753 , H486209 , H480374 , H481460 , H482897 , H484830 , H486291 , H480378 , H481461 , H483087 , H484842 , H486369 , H480379 , H481462 , H483204 , H484856 , H486385 , H480408 , H481483 , H483345 , H484950 , H486412 , H480409 , H481532 , H483350 , H484951 , H486442 , H480410 , H481533 , H483351 , H48SOOS , H486473 , H480422 , H481581 , H483352 , H485006 , H486486 , H480471 , H481582 , H483353 , H485090 , H486498 , H480477 , H481601 , H483354 , H485091 , H486499 , H480493 , H481627 , H483406 , H485113 , H486543 , H480566 , H481639 , H483442 , H485127 , H486579 , H4806Sl , H481685 , H483452 , H485143 , H486580 , H480692 , H481766 , H483592 , H485144 , H486615 , H480725 , H481781 , H483617 , H485145 , H486616 , H480764 , H481807 , H483666 , H485162 , H486688 , H480765 , H481815 , H483695 , H485244 , H486689 , H480766 , H481834 , H483696 , H485268 , H486694 , H480794 , H481857 , H483698 , H485308 , H486703 , H480808 , H481879 , H483723 , H485321 , H486716 , H480819 , H481923 , H483724 , H485322 , H486759 , H480820 , H481924 , H483725 , H485359 , H486760 , H480821 , H482166 , H483732 , H485442 , H486858 , H480917 , H482167 , H483733 , H485471 , H486859 , H480918 , H482168 , H483757 , H485481 , H486860 , H481054 , H482181 , H483758 , H485536 , H486861 , H481057 , H482182 , H483759 , H485537 , H486885 , H481112 , H482191 , H483767 , H485541 , H486895 , H481114 , H482192 , H483768 , H485557 , H486896 , H481119 , H482193 , H483840 , H485573 , H486938 , H481120 , H482194 , H483841 , H485580 , H486939 , H481133 , H482243 , H483871 , H485581 , H486941 , H481138 , H482244 , H483872 , H485582 , H486957 , H481154 , H482246 , H483943 , H485592 , H487011 , H481155 , H482266 , H483967 , H485595 , H487012 , H481164 , H482267 , H483968 , H485628 , H487277 , H481165 , H482323 , H483969 , H485689 , H487278 , H481166 , H482367 , H483993 , H485694 , H487279 , H481168 , H482385 , H484008 , H485706 , H487302 , H481171 , H482386 , H484023 , H485729 , H487428 , H481172 , H482387 , H484056 , H485776 , H487512 , H481196 , H482388 , H484153 , H485777 , H487513 , H481206 , H482391 , H484154 , H485792 , H487580 , H481207 , H482579 , H484168 , H485800 , H487621 , H481208 , H482580 , H484169 , H485890 , H487639 , H481211 , H482581 , H484274 , H485891 , H487698 , H481218 , H482638 , H484367 , H485892 , H487700 , H487771 , H48A323 , H48B875 , H48C805 , H487772 , H48A398 , H48B887 , H48C806 , H487775 , H48A462 , H48B898 , H48C807 , H487778 , H48A474 , H488899 , H48C808 , H487895 , H48A482 , H48B900 , H48C809 , H487950 , H48AS12 , H48C084 , H48C854 , H487980 , H48AS13 , H48C085 , H48C8SS , H487999 , H48ASSS , H48C086 , H48C866 , H488005 , H48AS87 , H48C116 , H48C881 , H488029 , H48AS90 , H48C120 , H4800Sl , H4880SS , H48AS91 , H48C121 , H480113 , H488102 , H48AS92 , H48C134 , H480118 , H488163 , H48AS93 , H48C135 , H480171 , H488210 , H48AS94 , H48C136 , H48D177 , H488211 , H48AS95 , H48C137 , H48D179 , H488229 , H48A612 , H48C138 , H48D187 , H488237 , H48A614 , H48C173 , H48D211 , H488238 , H48A713 , H48C174 , H48D235 , H488243 , H48A760 , H48C176 , H48D315 , H488249 , H48A762 , H48C212 , H48D316 , H488291 , H48A816 , H48C252 , H48D317 , H488311 , H48A818 , H48C253 , H48D318 , H488312 , H48A890 , H48C278 , H480319 , H488483 , H48A918 , H48C316 , H480320 , H489074 , H48A920 , H48C356 , H480321 , H489079 , H48A956 , H48C367 , H480347 , H489095 , H48A971 , H48C373 , H480349 , H489096 , H48A986 , H48C399 , H480353 , H489114 , H48A987 , H48C434 , H480354 , H48937. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The calculation of dose dynamic plans containing different energy fields is incorrect when the Pencil Beam Convolution 11.0 algorithm (PBC 11.0) is used.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Varian Medical Systems, Inc. Oncology Systems
- Distributor
- Worldwide Distribution - USA Nationwide
- Lot
- H480044 , H481353 , H482733 , H484585 , H485948 , H480045 , H481354 , H482734 , H484634 , H486024 , H480105 , H481433 , H482765 , H484673 , H486032 , H480106 , H481436 , H482827 , H484677 , H486093 , H480159 , H481437 , H482858 , H484680 , H486094 , H480160 , H481451 , H482859 , H484745 , H486095 , H480247 , H481452 , H482860 , H484746 , H486114 , H480301 , H481453 , H482888 , H484747 , H486140 , H480373 , H481454 , H482889 , H484753 , H486209 , H480374 , H481460 , H482897 , H484830 , H486291 , H480378 , H481461 , H483087 , H484842 , H486369 , H480379 , H481462 , H483204 , H484856 , H486385 , H480408 , H481483 , H483345 , H484950 , H486412 , H480409 , H481532 , H483350 , H484951 , H486442 , H480410 , H481533 , H483351 , H48SOOS , H486473 , H480422 , H481581 , H483352 , H485006 , H486486 , H480471 , H481582 , H483353 , H485090 , H486498 , H480477 , H481601 , H483354 , H485091 , H486499 , H480493 , H481627 , H483406 , H485113 , H486543 , H480566 , H481639 , H483442 , H485127 , H486579 , H4806Sl , H481685 , H483452 , H485143 , H486580 , H480692 , H481766 , H483592 , H485144 , H486615 , H480725 , H481781 , H483617 , H485145 , H486616 , H480764 , H481807 , H483666 , H485162 , H486688 , H480765 , H481815 , H483695 , H485244 , H486689 , H480766 , H481834 , H483696 , H485268 , H486694 , H480794 , H481857 , H483698 , H485308 , H486703 , H480808 , H481879 , H483723 , H485321 , H486716 , H480819 , H481923 , H483724 , H485322 , H486759 , H480820 , H481924 , H483725 , H485359 , H486760 , H480821 , H482166 , H483732 , H485442 , H486858 , H480917 , H482167 , H483733 , H485471 , H486859 , H480918 , H482168 , H483757 , H485481 , H486860 , H481054 , H482181 , H483758 , H485536 , H486861 , H481057 , H482182 , H483759 , H485537 , H486885 , H481112 , H482191 , H483767 , H485541 , H486895 , H481114 , H482192 , H483768 , H485557 , H486896 , H481119 , H482193 , H483840 , H485573 , H486938 , H481120 , H482194 , H483841 , H485580 , H486939 , H481133 , H482243 , H483871 , H485581 , H486941 , H481138 , H482244 , H483872 , H485582 , H486957 , H481154 , H482246 , H483943 , H485592 , H487011 , H481155 , H482266 , H483967 , H485595 , H487012 , H481164 , H482267 , H483968 , H485628 , H487277 , H481165 , H482323 , H483969 , H485689 , H487278 , H481166 , H482367 , H483993 , H485694 , H487279 , H481168 , H482385 , H484008 , H485706 , H487302 , H481171 , H482386 , H484023 , H485729 , H487428 , H481172 , H482387 , H484056 , H485776 , H487512 , H481196 , H482388 , H484153 , H485777 , H487513 , H481206 , H482391 , H484154 , H485792 , H487580 , H481207 , H482579 , H484168 , H485800 , H487621 , H481208 , H482580 , H484169 , H485890 , H487639 , H481211 , H482581 , H484274 , H485891 , H487698 , H481218 , H482638 , H484367 , H485892 , H487700 , H487771 , H48A323 , H48B875 , H48C805 , H487772 , H48A398 , H48B887 , H48C806 , H487775 , H48A462 , H48B898 , H48C807 , H487778 , H48A474 , H488899 , H48C808 , H487895 , H48A482 , H48B900 , H48C809 , H487950 , H48AS12 , H48C084 , H48C854 , H487980 , H48AS13 , H48C085 , H48C8SS , H487999 , H48ASSS , H48C086 , H48C866 , H488005 , H48AS87 , H48C116 , H48C881 , H488029 , H48AS90 , H48C120 , H4800Sl , H4880SS , H48AS91 , H48C121 , H480113 , H488102 , H48AS92 , H48C134 , H480118 , H488163 , H48AS93 , H48C135 , H480171 , H488210 , H48AS94 , H48C136 , H48D177 , H488211 , H48AS95 , H48C137 , H48D179 , H488229 , H48A612 , H48C138 , H48D187 , H488237 , H48A614 , H48C173 , H48D211 , H488238 , H48A713 , H48C174 , H48D235 , H488243 , H48A760 , H48C176 , H48D315 , H488249 , H48A762 , H48C212 , H48D316 , H488291 , H48A816 , H48C252 , H48D317 , H488311 , H48A818 , H48C253 , H48D318 , H488312 , H48A890 , H48C278 , H480319 , H488483 , H48A918 , H48C316 , H480320 , H489074 , H48A920 , H48C356 , H480321 , H489079 , H48A956 , H48C367 , H480347 , H489095 , H48A971 , H48C373 , H480349 , H489096 , H48A986 , H48C399 , H480353 , H489114 , H48A987 , H48C434 , H480354 , H48937
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVarian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29,May 22, 2013
- FDAVarian brand RPM Respiratory Gating System, RPM Gating v1.7.5 with 3D Option, Model Number: HS2 Product Usage:May 1, 2013
- FDAMerit Medical Systems Inc., Resolve Biliary Locking Drainage Catheter, Models RBC and RBDC, used for drainageFebruary 27, 2013
- FDAVascular Solutions, SuperCross Microcatheter, Sterilized using ethylene oxide, Rx Only, Model numbers 5300 -February 27, 2013
- FDAVarian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import ExportDecember 26, 2012
- FDAARIA Radiation Oncology, Version 11. Varian Medical Systems, Palo Alto, CA. The ARIA Radiation OncologyNovember 7, 2012
Questions about this recall
Was Varian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye. recalled?
Yes. FDA published recall Z-0867-2013 on February 27, 2013. Reported hazard: The calculation of dose dynamic plans containing different energy fields is incorrect when the Pencil Beam Convolution 11.0 algorithm (PBC 11.0) is used.
What should I do if I have this product?
Voluntary: Firm initiated · Letter