Back to search
FDAFDA · Food and drugs·Recall number Z-0870-2017

8470-009-030 3.0MM Precision Round, 16CM, single use, Sterile The Stryker MIS (Minimally Invasive Surgery) product offering, is used to remove osseous material during surgical cases. Each cutting accessory consists of: " A shank (B) with a notch (A), that fits in the attachment on one end and is connected to the head of the bur on the other end. " A head (C) connected to the shank (B) and removes material using machined flutes. The head is machined directly on the shank/neck of the bur.

Published December 28, 2016 · Recall initiated: November 30, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Stryker Instruments Div. of Stryker Corporation
Distributor
Worldwide Distribution - US (nationwide) including states of: AK, AL, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI and countries of: Canada, Japan, Venlo (Netherlands), France, and UK (United Kingdom).
Lot
Lot Number(s):16049017 & 16053017, Product Number: 8470-009-030
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card