Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (BIC) - VERSION 2, Article Code Number 3456102A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI 04046964285608), 710200U (UDI 04046964285615), 710200K (UDI 04046963686338), 710500L (UDI Not applicable) and 710500K (UDI Not applicable).
Published April 20, 2022 · Recall initiated: January 28, 2022
Summary
FDA records medical device Class II recall Z-0872-2022, published April 20, 2022 and initiated January 28, 2022. Recalling firm: B Braun Medical Inc; Status: Terminated; Reported quantity: 80; Lot: Lots 4/21 and 5/21. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- B Braun Medical Inc
- Distributor
- US Nationwide distribution.
- Lot
- Lots 4/21 and 5/21
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAB Braun SAFSITE Injection Site, RV1000NC SAFSITE VALVE W/O EXTRA CAP. Catalog Number: 415068.October 28, 2026
- FDABrand Name: Extension Set Product Name: DISCOFIX FOUR WAY STOPCOCK Model/Catalog Number: 456025September 9, 2026
- FDABrand Name: B. BRAUN MEDICAL INC. Product Name: CESK RIVERSIDE HEALTH SYSTEM Model/Catalog Number: 552126June 10, 2026
- FDABrand Name: B. BRAUN MEDICAL INC. Product Name: NES1725KFXL SPINAL EPID CLOSED TIP CATH Model/Catalog Number:June 10, 2026
- FDAGiant Eagle, Happy Face Cookie 4 Pack 25051900000, 11 oz, Clam ShellApril 20, 2022
- FDAConductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSIONApril 20, 2022
Questions about this recall
Was Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (BIC) - VERSION 2, Article Code Number 3456102A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI 04046964285608), 710200U (UDI 04046964285615), 710200K (UDI 04046963686338), 710500L (UDI Not applicable) and 710500K (UDI Not applicable). recalled?
Yes. FDA published recall Z-0872-2022 on April 20, 2022. Reported hazard: Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.
What should I do if I have this product?
Voluntary: Firm initiated · Letter