Intera 1.0T Power/Pulsar, Model Number 781103. Nuclear Magnetic Resonance Imaging System

Published March 6, 2019 · Recall initiated: December 21, 2018

Summary

FDA records medical device Class II recall Z-0876-2019, published March 6, 2019 and initiated December 21, 2018. Recalling firm: Philips Medical Systems Nederlands; Status: Terminated; Reported quantity: 11226 total; Lot: All systems. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Medical Systems Nederlands
Distributor
U.S. Nationwide distribution.
Lot
All systems
Risk classification
Class II
Area (FDA)
device
Country
Netherlands

Related recalls

Questions about this recall

Was Intera 1.0T Power/Pulsar, Model Number 781103. Nuclear Magnetic Resonance Imaging System recalled?

Yes. FDA published recall Z-0876-2019 on March 6, 2019. Reported hazard: There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.

What should I do if I have this product?

Voluntary: Firm initiated · Letter