Intera 1.0T Power/Pulsar, Model Number 781103. Nuclear Magnetic Resonance Imaging System
Published March 6, 2019 · Recall initiated: December 21, 2018
Summary
FDA records medical device Class II recall Z-0876-2019, published March 6, 2019 and initiated December 21, 2018. Recalling firm: Philips Medical Systems Nederlands; Status: Terminated; Reported quantity: 11226 total; Lot: All systems. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems Nederlands
- Distributor
- U.S. Nationwide distribution.
- Lot
- All systems
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands
Related recalls
- FDAAchieva XR 1.5T/3.0T, Model Number 781153. Nuclear Magnetic Resonance Imaging SystemMarch 6, 2019
- FDASmartPath to dStream for XR and 3.0T, Model Number 781270. Nuclear Magnetic Resonance Imaging SystemMarch 6, 2019
- FDAIntera 1.5T Achieva CV Nova, Model Number 781174. Nuclear Magnetic Resonance Imaging SystemMarch 6, 2019
- FDAIntera Achieva 1.5T Nova Dual, Model Number 781173. Nuclear Magnetic Resonance Imaging SystemMarch 6, 2019
- FDALIFEPAK 15 Monitor/Defibrillator Product Usage: The LIFEPAK(R) 15 Monitor/defibrillator (LP15) is a completeMarch 6, 2019
- FDAEnterprise 1.5T, Model Number 781145. Nuclear Magnetic Resonance Imaging SystemMarch 6, 2019
Questions about this recall
Was Intera 1.0T Power/Pulsar, Model Number 781103. Nuclear Magnetic Resonance Imaging System recalled?
Yes. FDA published recall Z-0876-2019 on March 6, 2019. Reported hazard: There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
What should I do if I have this product?
Voluntary: Firm initiated · Letter