Back to search
FDAFDA · Food and drugs·Recall number Z-0879-2020

2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERILE, Part Number 04.210.110TS - Product Usage:Sterile Tube Packaging is a packaging system to deliver single, sterile screws to the sterile field.

Published February 5, 2020 · Recall initiated: December 16, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Synthes, Inc.
Distributor
US Nationwide distribution in the states of AL and ID.
Lot
Lot Number: 6L37877
Risk classification
Class II
Area (FDA)
device
Country
United States