FDAFDA · Food and drugs·Recall number Z-0893-2013

Stryker XIA 3 Iliac Screwdriver, 3-PCS. The XIA 3 Iliac Screwdriver is used to drive XIA 3 closed head Monoaxial and closed head Polyaxial screws into the Ilium. Orthopedic manual surgical instrument.

Published March 6, 2013 · Recall initiated: December 27, 2012

Summary

FDA records medical device Class II recall Z-0893-2013, published March 6, 2013 and initiated December 27, 2012. Recalling firm: Stryker Spine; Status: Terminated; Reported quantity: 46 units US; Lot: Catalog Number: 48231326. Lot Number: 10G637; 10G638 and 10E852. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Surgeons have experienced unthreading of the screwdrivers outer shafts during use.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Stryker Spine
Distributor
Nationwide Distribution including the states of AL, AZ, CA, CO, FL, GA, IA, IL,IN, KS, LA, MA, MD, MI, MO, MN, NC, NY. OH, OK, OR, PA, TN, TX, UT and WI.
Lot
Catalog Number: 48231326. Lot Number: 10G637; 10G638 and 10E852.
Risk classification
Class II
Area (FDA)
device
Country
United States