SmartPath to dStream for 1.5T, Model Number 781260. Nuclear Magnetic Resonance Imaging System
Published March 6, 2019 · Recall initiated: December 21, 2018
Summary
FDA records medical device Class II recall Z-0900-2019, published March 6, 2019 and initiated December 21, 2018. Recalling firm: Philips Medical Systems Nederlands; Status: Terminated; Reported quantity: 11226 total; Lot: All systems. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems Nederlands
- Distributor
- U.S. Nationwide distribution.
- Lot
- All systems
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands
Related recalls
- FDAAchieva XR 1.5T/3.0T, Model Number 781153. Nuclear Magnetic Resonance Imaging SystemMarch 6, 2019
- FDASmartPath to dStream for XR and 3.0T, Model Number 781270. Nuclear Magnetic Resonance Imaging SystemMarch 6, 2019
- FDAIntera 1.5T Achieva CV Nova, Model Number 781174. Nuclear Magnetic Resonance Imaging SystemMarch 6, 2019
- FDAIntera Achieva 1.5T Nova Dual, Model Number 781173. Nuclear Magnetic Resonance Imaging SystemMarch 6, 2019
- FDAAchieva 1.5T, Model Number 781296. Nuclear Magnetic Resonance Imaging SystemMarch 6, 2019
- FDAIngenia 3.0T, Model Number 781342. Nuclear Magnetic Resonance Imaging SystemMarch 6, 2019
Questions about this recall
Was SmartPath to dStream for 1.5T, Model Number 781260. Nuclear Magnetic Resonance Imaging System recalled?
Yes. FDA published recall Z-0900-2019 on March 6, 2019. Reported hazard: There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
What should I do if I have this product?
Voluntary: Firm initiated · Letter