Medtronic Advanced Energy Aquamantys3 Pump Generator; Product Catalog Number: 10-1357 (Generator assembly), 40-404-1(as shipped, including accessories) The Aquamantys3 System combines radio frequency (RF) energy and saline to reduce blood loss during and after surgical procedures. This patented Transcollation technology has been shown to reduce transfusion rates and may also reduce the need for other blood management procedures.
Published February 12, 2014 · Recall initiated: November 12, 2013
Summary
FDA records medical device Class II recall Z-0905-2014, published February 12, 2014 and initiated November 12, 2013. Recalling firm: Medtronic Advanced Energy, LLC; Status: Terminated; Reported quantity: 242 devices; Lot: lowest serial number is GN001141; the highest is GN001704 (non-consecutive). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The internal protection circuitry on the electrosurgical output that prevents a patient from exposure to DC voltage when the device is activated is being compromised during normal use.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Advanced Energy, LLC
- Distributor
- Nationwide Distribution including CA, MO, PA, NJ, NY, OH, VA, TX, NC, DE, KS, TN, LA, IL, MI, SC, WI, and FL.
- Lot
- lowest serial number is GN001141; the highest is GN001704 (non-consecutive)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntact Disposable Wands (part of the Intact System) 777-115: 5 Pack 15mm Disposable Product Usage: The IntactFebruary 8, 2017
- FDAIntact Disposable Wands (part of the Intact System) 777-112BT: 5 Pack 12mm Disposable Product Usage: TheFebruary 8, 2017
- FDAIntact Disposable Wands (part of the Intact System) 777-115BT: 5 Pack 15mm Disposable Product Usage: TheFebruary 8, 2017
- FDAIntact Disposable Wands (part of the Intact System) 900-115: 5 Pack 15mm Disposable Product Usage: The IntactFebruary 8, 2017
- FDAABX PENTRA N Control, Model No. A11A01653 The ABX PENTRA N Control is for use in quality control by monitoringFebruary 12, 2014
- FDA***REF 186-01-08***InteGrip ACETABULAR AUGMENT***SMALL 8mm. Use with 48/50mm Shell.***EXACTECH, Gainesville,February 12, 2014
Questions about this recall
Was Medtronic Advanced Energy Aquamantys3 Pump Generator; Product Catalog Number: 10-1357 (Generator assembly), 40-404-1(as shipped, including accessories) The Aquamantys3 System combines radio frequency (RF) energy and saline to reduce blood loss during and after surgical procedures. This patented Transcollation technology has been shown to reduce transfusion rates and may also reduce the need for other blood management procedures. recalled?
Yes. FDA published recall Z-0905-2014 on February 12, 2014. Reported hazard: The internal protection circuitry on the electrosurgical output that prevents a patient from exposure to DC voltage when the device is activated is being compromised during normal use.
What should I do if I have this product?
Voluntary: Firm initiated · Letter