Olympus, Model No. CF-Q140L, flexible scope, large diameter, video colonoscope

Published March 2, 2016 · Recall initiated: January 20, 2016

Summary

FDA records medical device Class II recall Z-0909-2016, published March 2, 2016 and initiated January 20, 2016. Recalling firm: Integrated Medical Systems Inc; Status: Terminated; Reported quantity: 2 instruments; Lot: Serial Numbers: 2913081,2600279. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Integrated Medical Systems Inc
Distributor
Distributed US (nationwide) and in Australia.
Lot
Serial Numbers: 2913081,2600279
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Olympus, Model No. CF-Q140L, flexible scope, large diameter, video colonoscope recalled?

Yes. FDA published recall Z-0909-2016 on March 2, 2016. Reported hazard: If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.

What should I do if I have this product?

Voluntary: Firm initiated · Letter