FDAFDA · Food and drugs·Recall number Z-0918-2014

Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation.

Published February 12, 2014 · Recall initiated: January 20, 2014

Summary

FDA records medical device Class II recall Z-0918-2014, published February 12, 2014 and initiated January 20, 2014. Recalling firm: Edwards Lifesciences, LLC; Status: Terminated; Reported quantity: 422; Lot: lot 59617777 and 59658495. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open position. As a result, user may have difficulty fitting a 26mm Sapien transcatheter heart valve into the crimper orifice when preparing the valve for crimping.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Edwards Lifesciences, LLC
Distributor
Nationwide Distribution
Lot
lot 59617777 and 59658495.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation. recalled?

Yes. FDA published recall Z-0918-2014 on February 12, 2014. Reported hazard: Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open position. As a result, user may have difficulty fitting a 26mm Sapien transcatheter heart valve into the crimper orifice when preparing the valve for crimping.

What should I do if I have this product?

Voluntary: Firm initiated · Letter