Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation.
Published February 12, 2014 · Recall initiated: January 20, 2014
Summary
FDA records medical device Class II recall Z-0918-2014, published February 12, 2014 and initiated January 20, 2014. Recalling firm: Edwards Lifesciences, LLC; Status: Terminated; Reported quantity: 422; Lot: lot 59617777 and 59658495. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open position. As a result, user may have difficulty fitting a 26mm Sapien transcatheter heart valve into the crimper orifice when preparing the valve for crimping.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Edwards Lifesciences, LLC
- Distributor
- Nationwide Distribution
- Lot
- lot 59617777 and 59658495.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGeneric Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery SystemNovember 11, 2026
- FDAEdwards EVOQUE tricuspid delivery system REF:9850TDS Sterile EOMay 20, 2026
- FDABrand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016AS SoftwareJuly 16, 2025
- FDABrand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI16July 16, 2025
- FDAUpgrade Kit, Resuscitation, T-Piece, part numbers M1139417, M1192226 and M1223628, labeled in part ***GEFebruary 12, 2014
- FDAIMPAX RIS QDOC 5.8February 12, 2014
Questions about this recall
Was Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation. recalled?
Yes. FDA published recall Z-0918-2014 on February 12, 2014. Reported hazard: Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open position. As a result, user may have difficulty fitting a 26mm Sapien transcatheter heart valve into the crimper orifice when preparing the valve for crimping.
What should I do if I have this product?
Voluntary: Firm initiated · Letter