Cable remote control JUPITER / TS7500, Material no.1601926, Serial number: to (incl.) 101465919. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.
Published January 14, 2015 · Recall initiated: December 3, 2014
Summary
FDA records medical device Class II recall Z-0919-2015, published January 14, 2015 and initiated December 3, 2014. Recalling firm: Trumpf Medical Systems, Inc.; Status: Terminated; Reported quantity: 3303 total all devices; Lot: The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Trumpf Medical Systems, Inc.
- Distributor
- Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.
- Lot
- The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFloor Mounting Column TS 7500 U, Material Number 1730732; Floor Mounting Column ARTIS, Material Number 1501878October 12, 2016
- FDAMobile Column TruSystem 7500, Material Number 1717023; Mobile Column TruSystem 5500, Material Number 1607822October 12, 2016
- FDAOperating Table TruSystem 7000 U, Material Number 1604788; Operating Table TruSystgem 7000 U (MB), MaterialOctober 12, 2016
- FDAStationary Column TruSystem 7599, Material Number 1717020 Patient positioning during surgery, ranging fromOctober 12, 2016
- FDAe.cam -emission computed tomography system Used to detect or image the distribution of radionuclides in theJanuary 14, 2015
- FDAIR remote control JUPITER U, Material no. 4500120, Serial number: to (incl.) 102306940. Patient positioningJanuary 14, 2015
Questions about this recall
Was Cable remote control JUPITER / TS7500, Material no.1601926, Serial number: to (incl.) 101465919. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system. recalled?
Yes. FDA published recall Z-0919-2015 on January 14, 2015. Reported hazard: Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
What should I do if I have this product?
Voluntary: Firm initiated · Letter