FDAFDA · Food and drugs·Recall number Z-0919-2015

Cable remote control JUPITER / TS7500, Material no.1601926, Serial number: to (incl.) 101465919. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Published January 14, 2015 · Recall initiated: December 3, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Trumpf Medical Systems, Inc.
Distributor
Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.
Lot
The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.
Risk classification
Class II
Area (FDA)
device
Country
United States