SOMATOM Perspective, Computed tomography x-ray system
Published January 4, 2017 · Recall initiated: November 22, 2016
Summary
FDA records medical device Class II recall Z-0921-2017, published January 4, 2017 and initiated November 22, 2016. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 13 systems; Lot: Model # 10967888 Serial Number:59376 91766 91767 115456 115454 59380 77560 98840 77557 91758 59377 59378 115451. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- Distributed to: IL IL GA SC SD OK NY MI IN MD AR
- Lot
- Model # 10967888 Serial Number:59376 91766 91767 115456 115454 59380 77560 98840 77557 91758 59377 59378 115451
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDAShimadzu X-ray High Voltage Generator Model No. UD150B-40 on SONIALVISION safire II/17 systems withJanuary 4, 2017
- FDAArthex S3 Arm Sleeve Connector; Catalog Number AR-1650-02; manufactured and distributed between March andJanuary 4, 2017
Questions about this recall
Was SOMATOM Perspective, Computed tomography x-ray system recalled?
Yes. FDA published recall Z-0921-2017 on January 4, 2017. Reported hazard: Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.
What should I do if I have this product?
Voluntary: Firm initiated · Letter