FDAFDA · Food and drugs·Recall number Z-0925-2013

REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument

Published March 20, 2013 · Recall initiated: February 5, 2013

Summary

FDA records medical device Class II recall Z-0925-2013, published March 20, 2013 and initiated February 5, 2013. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 58; Lot: Catalog Number 5313152 Lot 153880,153930, 415500, 648010. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Biomet, Inc.
Distributor
Worldwide Distribution - No USA distribution -Internationally to Canada, Japan, and Chile
Lot
Catalog Number 5313152 Lot 153880,153930, 415500, 648010
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument recalled?

Yes. FDA published recall Z-0925-2013 on March 20, 2013. Reported hazard: The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit