REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument
Published March 20, 2013 · Recall initiated: February 5, 2013
Summary
FDA records medical device Class II recall Z-0925-2013, published March 20, 2013 and initiated February 5, 2013. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 58; Lot: Catalog Number 5313152 Lot 153880,153930, 415500, 648010. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomet, Inc.
- Distributor
- Worldwide Distribution - No USA distribution -Internationally to Canada, Japan, and Chile
- Lot
- Catalog Number 5313152 Lot 153880,153930, 415500, 648010
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475640 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475625 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475630 Product Description: Intramedullary BoneAugust 12, 2026
- FDAGenesis Straight Esthetic 5.0 mm 3. mm Cuff Abutment Part Number: G71205 Genesis implant for single stage orMarch 20, 2013
- FDAMedtronic O-arm Imaging System Product Usage: image-intensified fluoroscopic x-ray system, mobileMarch 20, 2013
Questions about this recall
Was REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument recalled?
Yes. FDA published recall Z-0925-2013 on March 20, 2013. Reported hazard: The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit