REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument

Published March 20, 2013 · Recall initiated: February 5, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Biomet, Inc.
Distributor
Worldwide Distribution - No USA distribution -Internationally to Canada, Japan, and Chile
Lot
Catalog Number 5313152 Lot 153880,153930, 415500, 648010
Risk classification
Class II
Area (FDA)
device
Country
United States