BrightView XCT, gamma camera for SPECT; Model Nos.: 6-digit format 882482; 12-digit format 453560462131 453560749161

Published February 14, 2024 · Recall initiated: December 15, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips North America
Distributor
Domestic distribution nationwide. International distribution worldwide.
Lot
All serial numbers are affected. No UDI-DI available for this product.
Risk classification
Class I
Area (FDA)
device
Country
United States