Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of the IMRIS HFD100 head fixation device, models #119629-000 and #119630-000, which are used with ORT400 operating room tables.

Published April 20, 2022 · Recall initiated: February 17, 2022

Summary

FDA records medical device Class II recall Z-0934-2022, published April 20, 2022 and initiated February 17, 2022. Recalling firm: Deerfield Imaging, Inc.; Status: Terminated; Reported quantity: 24; Lot: Model #119629-000, UDI 00857534006592; and model #119630-000, UDI 00857534006615. Serial numbers: 10008164, 10007550, 10007555, 10008864, 10008160, 10008161, 10008859, 10008860, 10010668, 10010669, 10010673, 10010674, 10010672, 10007556, 10007557, 10008862, 10008861, 10008158, 10008159, 10008863, 10008165, 10007558, 10008865, and 10010675. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may deflect (rotationally around the axis) and not return to within acceptable distance from its prior position after the force is removed. In order to reduce deflection at the table mount location, users have tightened the hand knob screw to a degree that it is then difficult to untighten it later in the clinical workflow. Rotational movement at the table attachment point could result in HFD movement going unnoticed, which could result in a situation where surgical or medical intervention is necessary, a situation where navigation needs to be reestablished, or a situation resulting in extended time under anesthesia for the patient. It may also be difficult to loosen the mount clamp component after use, potentially leading to ergonomic injury or a workaround during equipment teardown.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Deerfield Imaging, Inc.
Distributor
Worldwide distribution - US Nationwide distribution in the states of AL, FL, GA, MA, MN, NC, NY, OH, and TX. The countries of Sweden and Switzerland.
Lot
Model #119629-000, UDI 00857534006592; and model #119630-000, UDI 00857534006615. Serial numbers: 10008164, 10007550, 10007555, 10008864, 10008160, 10008161, 10008859, 10008860, 10010668, 10010669, 10010673, 10010674, 10010672, 10007556, 10007557, 10008862, 10008861, 10008158, 10008159, 10008863, 10008165, 10007558, 10008865, and 10010675.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of the IMRIS HFD100 head fixation device, models #119629-000 and #119630-000, which are used with ORT400 operating room tables. recalled?

Yes. FDA published recall Z-0934-2022 on April 20, 2022. Reported hazard: Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may deflect (rotationally around the axis) and not return to within acceptable distance from its prior position after the force is removed. In order to reduce deflection at the table mount location, users have tightened the hand knob screw to a degree that it is then difficult to untighten it later in the clinical workflow. Rotational movement at the table attachment point could result in HFD movement going unnoticed, which could result in a situation where surgical or medical intervention is necessary, a situation where navigation needs to be reestablished, or a situation resulting in extended time under anesthesia for the patient. It may also be difficult to loosen the mount clamp component after use, potentially leading to ergonomic injury or a workaround during equipment teardown.

What should I do if I have this product?

Voluntary: Firm initiated · Letter