Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of the IMRIS HFD100 head fixation device, models #119629-000 and #119630-000, which are used with ORT400 operating room tables.
Published April 20, 2022 · Recall initiated: February 17, 2022
Summary
FDA records medical device Class II recall Z-0934-2022, published April 20, 2022 and initiated February 17, 2022. Recalling firm: Deerfield Imaging, Inc.; Status: Terminated; Reported quantity: 24; Lot: Model #119629-000, UDI 00857534006592; and model #119630-000, UDI 00857534006615. Serial numbers: 10008164, 10007550, 10007555, 10008864, 10008160, 10008161, 10008859, 10008860, 10010668, 10010669, 10010673, 10010674, 10010672, 10007556, 10007557, 10008862, 10008861, 10008158, 10008159, 10008863, 10008165, 10007558, 10008865, and 10010675. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may deflect (rotationally around the axis) and not return to within acceptable distance from its prior position after the force is removed. In order to reduce deflection at the table mount location, users have tightened the hand knob screw to a degree that it is then difficult to untighten it later in the clinical workflow. Rotational movement at the table attachment point could result in HFD movement going unnoticed, which could result in a situation where surgical or medical intervention is necessary, a situation where navigation needs to be reestablished, or a situation resulting in extended time under anesthesia for the patient. It may also be difficult to loosen the mount clamp component after use, potentially leading to ergonomic injury or a workaround during equipment teardown.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Deerfield Imaging, Inc.
- Distributor
- Worldwide distribution - US Nationwide distribution in the states of AL, FL, GA, MA, MN, NC, NY, OH, and TX. The countries of Sweden and Switzerland.
- Lot
- Model #119629-000, UDI 00857534006592; and model #119630-000, UDI 00857534006615. Serial numbers: 10008164, 10007550, 10007555, 10008864, 10008160, 10008161, 10008859, 10008860, 10010668, 10010669, 10010673, 10010674, 10010672, 10007556, 10007557, 10008862, 10008861, 10008158, 10008159, 10008863, 10008165, 10007558, 10008865, and 10010675.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADiclofenac Sodium Topical Solution 1.5% w/w, Generic for Pennsaid, 150 ml bottle, Rx Only, Mfg: SolaApril 27, 2022
- FDAThe IFU for the Autoclavable Temperature Probes: Part # 453564635891, providing in the packaging for theApril 20, 2022
- FDAPlus 30 Priority Pack w/.096 RHV - REF 1000183 WPL2122277-01 (20-03-01) Plus 30 Priority Pack w/.115 RHV - REFApril 20, 2022
- FDAIMRIS MR/X-ray head fixation device, Model HFD200October 30, 2019
- FDAIMRIS T2X table, Part Numbers: (a)108000-000, (b)108000-600, (c)109682-000, (d) 109682-600, (e) 110470-600 -December 26, 2018
- FDAIMRIS ORT300, Removable Operating Room Table, Part Numbers: (a) 114093-000 (b) 114093-600April 4, 2018
Questions about this recall
Was Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of the IMRIS HFD100 head fixation device, models #119629-000 and #119630-000, which are used with ORT400 operating room tables. recalled?
Yes. FDA published recall Z-0934-2022 on April 20, 2022. Reported hazard: Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may deflect (rotationally around the axis) and not return to within acceptable distance from its prior position after the force is removed. In order to reduce deflection at the table mount location, users have tightened the hand knob screw to a degree that it is then difficult to untighten it later in the clinical workflow. Rotational movement at the table attachment point could result in HFD movement going unnoticed, which could result in a situation where surgical or medical intervention is necessary, a situation where navigation needs to be reestablished, or a situation resulting in extended time under anesthesia for the patient. It may also be difficult to loosen the mount clamp component after use, potentially leading to ergonomic injury or a workaround during equipment teardown.
What should I do if I have this product?
Voluntary: Firm initiated · Letter