Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.
Published March 20, 2013 · Recall initiated: December 13, 2012
Summary
FDA records medical device Class II recall Z-0939-2013, published March 20, 2013 and initiated December 13, 2012. Recalling firm: Draeger Medical, Inc.; Status: Terminated; Reported quantity: 260; Lot: Catalog Number MS20783 (Infinity MCable-Dual Hemo). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Draeger Medical, Inc.
- Distributor
- Distributed in the states of CA, IL, and MA.
- Lot
- Catalog Number MS20783 (Infinity MCable-Dual Hemo)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesiaDecember 25, 2024
- FDABrand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesiaDecember 25, 2024
- FDABrand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicableNovember 20, 2024
- FDABrand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not ApplicableNovember 20, 2024
- FDASynthes Spine 5.2mm Titanium Click ' X Pedicle Screw Dual Core 40 mm Thread Length, Part Number 487.072.March 20, 2013
- FDAGOPump Elastometric infusion PumpKit with/ and without accesssories all containing Dual GOPump, part numberMarch 20, 2013
Questions about this recall
Was Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system. recalled?
Yes. FDA published recall Z-0939-2013 on March 20, 2013. Reported hazard: The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.
What should I do if I have this product?
Voluntary: Firm initiated · Letter