Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.

Published March 20, 2013 · Recall initiated: December 13, 2012

Summary

FDA records medical device Class II recall Z-0939-2013, published March 20, 2013 and initiated December 13, 2012. Recalling firm: Draeger Medical, Inc.; Status: Terminated; Reported quantity: 260; Lot: Catalog Number MS20783 (Infinity MCable-Dual Hemo). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Draeger Medical, Inc.
Distributor
Distributed in the states of CA, IL, and MA.
Lot
Catalog Number MS20783 (Infinity MCable-Dual Hemo)
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system. recalled?

Yes. FDA published recall Z-0939-2013 on March 20, 2013. Reported hazard: The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.

What should I do if I have this product?

Voluntary: Firm initiated · Letter