Back to search
FDAFDA · Food and drugs·Recall number Z-0940-2017

TVC Imaging System model TVC-MC9 , for the near-infrared examination of coronary arteries in patients undergoing invasive coronary angiography The TVC Imaging System is an intravascular imaging device that simultaneously assesses vessel composition and structure using near-infrared spectroscopy (NIRS) and intravascular ultrasound (IVUS). This dual-modality instrument performs near-infrared spectroscopic analysis of the vessel to detect lipid core-containing plaques of interest (LCP) displayed in a map called a Chemogram, and simultaneously generates high resolution IVUS images that display structural details of the vessel and plaque in transverse and longitudinal views

Published January 11, 2017 · Recall initiated: December 15, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Near-infrared spectroscopic image (Chemogram) generated by the TVC Imaging System TVC-MC9 in some cases demonstrated an unusual appearance in that the yellow in the image (indicating the presence of lipid core plaques) was horizontally-oriented

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Infraredx, Inc.
Distributor
Worldwide Distribution - US including MI and Internationally to South Korea
Lot
Serial Numbers: MC9-15013, MC9-15006, MC9-15005, MC9-15008, MC9-15012, MC9-15007 and 12718M00009 (spare laser assembly)
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card