TVC Imaging System model TVC-MC9 , for the near-infrared examination of coronary arteries in patients undergoing invasive coronary angiography The TVC Imaging System is an intravascular imaging device that simultaneously assesses vessel composition and structure using near-infrared spectroscopy (NIRS) and intravascular ultrasound (IVUS). This dual-modality instrument performs near-infrared spectroscopic analysis of the vessel to detect lipid core-containing plaques of interest (LCP) displayed in a map called a Chemogram, and simultaneously generates high resolution IVUS images that display structural details of the vessel and plaque in transverse and longitudinal views

Published January 11, 2017 · Recall initiated: December 15, 2016

Summary

FDA records medical device Class II recall Z-0940-2017, published January 11, 2017 and initiated December 15, 2016. Recalling firm: Infraredx, Inc.; Status: Terminated; Reported quantity: 7 units; Lot: Serial Numbers: MC9-15013, MC9-15006, MC9-15005, MC9-15008, MC9-15012, MC9-15007 and 12718M00009 (spare laser assembly). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Near-infrared spectroscopic image (Chemogram) generated by the TVC Imaging System TVC-MC9 in some cases demonstrated an unusual appearance in that the yellow in the image (indicating the presence of lipid core plaques) was horizontally-oriented

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Infraredx, Inc.
Distributor
Worldwide Distribution - US including MI and Internationally to South Korea
Lot
Serial Numbers: MC9-15013, MC9-15006, MC9-15005, MC9-15008, MC9-15012, MC9-15007 and 12718M00009 (spare laser assembly)
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was TVC Imaging System model TVC-MC9 , for the near-infrared examination of coronary arteries in patients undergoing invasive coronary angiography The TVC Imaging System is an intravascular imaging device that simultaneously assesses vessel composition and structure using near-infrared spectroscopy (NIRS) and intravascular ultrasound (IVUS). This dual-modality instrument performs near-infrared spectroscopic analysis of the vessel to detect lipid core-containing plaques of interest (LCP) displayed in a map called a Chemogram, and simultaneously generates high resolution IVUS images that display structural details of the vessel and plaque in transverse and longitudinal views recalled?

Yes. FDA published recall Z-0940-2017 on January 11, 2017. Reported hazard: Near-infrared spectroscopic image (Chemogram) generated by the TVC Imaging System TVC-MC9 in some cases demonstrated an unusual appearance in that the yellow in the image (indicating the presence of lipid core plaques) was horizontally-oriented

What should I do if I have this product?

Voluntary: Firm initiated · Letter