Ingenia 1.5T CX. Model (REF) Numbers 781262, 781261.
Published February 7, 2024 · Recall initiated: January 3, 2024
Summary
FDA records medical device Class II recall Z-0950-2024, published February 7, 2024 and initiated January 3, 2024. Recalling firm: Philips North America; Status: Ongoing; Reported quantity: 0 US; 146 OUS; Lot: a. Model No. 781262; UDI-DI 00884838068445. b. Model No. 781261; UDI-DI 00884838068438. All units affected. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America
- Distributor
- Worldwide - US Nationwide distribution.
- Lot
- a. Model No. 781262; UDI-DI 00884838068445. b. Model No. 781261; UDI-DI 00884838068438. All units affected
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAModel Number: 712010.November 4, 2026
- FDAPhilips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;October 28, 2026
- FDAIngenia Ambition S and Ingenia Ambition X.; Model number (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S),October 7, 2026
- FDASmartPath to MR 7700; Model number (REF): 1. 782178, 2. 782207; Software Versions: R11, R12 and R13;October 7, 2026
- FDAAchieva XR. Model (REF) Numbers 781153, 781253.February 7, 2024
- FDAIngenia 3.0T CX. Model (REF) Numbers 781271, 782105.February 7, 2024
Questions about this recall
Was Ingenia 1.5T CX. Model (REF) Numbers 781262, 781261. recalled?
Yes. FDA published recall Z-0950-2024 on February 7, 2024. Reported hazard: The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
What should I do if I have this product?
Voluntary: Firm initiated · Letter