Back to search
FDAFDA · Food and drugs·Recall number Z-0955-2024

SmartPath to dStream for 1.5T. Model (REF) Numbers 781260, 782112.

Published February 7, 2024 · Recall initiated: January 3, 2024

Summary

FDA records medical device Class II recall Z-0955-2024, published February 7, 2024 and initiated January 3, 2024. Recalling firm: Philips North America; Status: Ongoing; Reported quantity: 37 US; 735 OUS; Lot: a. Model No. 781260; UDI-DI 00884838095076. b. Model No. 782112; UDI-DI 00884838098886. All units affected. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips North America
Distributor
Worldwide - US Nationwide distribution.
Lot
a. Model No. 781260; UDI-DI 00884838095076. b. Model No. 782112; UDI-DI 00884838098886. All units affected
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SmartPath to dStream for 1.5T. Model (REF) Numbers 781260, 782112. recalled?

Yes. FDA published recall Z-0955-2024 on February 7, 2024. Reported hazard: The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.

What should I do if I have this product?

Voluntary: Firm initiated · Letter