Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve DD, Model A-TCSE-DD, Sterile.
Published April 27, 2022 · Recall initiated: March 24, 2022
Summary
FDA records medical device Class II recall Z-0957-2022, published April 27, 2022 and initiated March 24, 2022. Recalling firm: Abbott; Status: Terminated; Reported quantity: 588 devices; Lot: Lot numbers: 8181068, 8182131, 8198290, 8198350, 8200249, 8200300, 8201447, 8202313, 8203518, 8203835, 8203857, 8205760, and 8207547. UDI (01)05415067027634(17)230331(10)xxxxxxx (lot number). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Devices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott
- Distributor
- Worldwide distribution - US Nationwide - there was government distribution but no military. The countries of Austria, Belgium, Canada, Denmark, France, Germany, Greece, Israel, Italy, Japan, Netherlands, Norway, Palestine, Poland, Portugal, Russia, Slovakia, Spain, Sweden, Taiwan, and United Kingdom.
- Lot
- Lot numbers: 8181068, 8182131, 8198290, 8198350, 8200249, 8200300, 8201447, 8202313, 8203518, 8203835, 8203857, 8205760, and 8207547. UDI (01)05415067027634(17)230331(10)xxxxxxx (lot number).
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAmplatzer Amulet Left Atrial Appendage Occluder Model Numbers 1. 9-ACP2-007-016 2. 9-ACP2-007-018 3.October 28, 2026
- FDATactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, FJ, (BiD Curve F-J), REF A-TFSE-FJ; cardiac catheterOctober 22, 2025
- FDATactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, DD, (BiD Curve D-D), REF A-TFSE-DD; cardiac catheterOctober 22, 2025
- FDATactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, D, (UniD Curve D), REF A-TFSE-D; cardiac catheterOctober 22, 2025
- FDAInactivated macrolide and quinolone-resistant Mycoplasma genitaliumApril 27, 2022
- FDAArtis zee multi-purpose, Model Number 10094139April 27, 2022
Questions about this recall
Was Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve DD, Model A-TCSE-DD, Sterile. recalled?
Yes. FDA published recall Z-0957-2022 on April 27, 2022. Reported hazard: Devices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device.
What should I do if I have this product?
Voluntary: Firm initiated · Letter