Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve DD, Model A-TCSE-DD, Sterile.

Published April 27, 2022 · Recall initiated: March 24, 2022

Summary

FDA records medical device Class II recall Z-0957-2022, published April 27, 2022 and initiated March 24, 2022. Recalling firm: Abbott; Status: Terminated; Reported quantity: 588 devices; Lot: Lot numbers: 8181068, 8182131, 8198290, 8198350, 8200249, 8200300, 8201447, 8202313, 8203518, 8203835, 8203857, 8205760, and 8207547. UDI (01)05415067027634(17)230331(10)xxxxxxx (lot number). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Devices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Abbott
Distributor
Worldwide distribution - US Nationwide - there was government distribution but no military. The countries of Austria, Belgium, Canada, Denmark, France, Germany, Greece, Israel, Italy, Japan, Netherlands, Norway, Palestine, Poland, Portugal, Russia, Slovakia, Spain, Sweden, Taiwan, and United Kingdom.
Lot
Lot numbers: 8181068, 8182131, 8198290, 8198350, 8200249, 8200300, 8201447, 8202313, 8203518, 8203835, 8203857, 8205760, and 8207547. UDI (01)05415067027634(17)230331(10)xxxxxxx (lot number).
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve DD, Model A-TCSE-DD, Sterile. recalled?

Yes. FDA published recall Z-0957-2022 on April 27, 2022. Reported hazard: Devices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device.

What should I do if I have this product?

Voluntary: Firm initiated · Letter