Artis zee biplane MN, Model Number 10094143
Published April 27, 2022 · Recall initiated: February 23, 2022
Summary
FDA records medical device Class II recall Z-0972-2022, published April 27, 2022 and initiated February 23, 2022. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Ongoing; Reported quantity: 7 systems; Lot: UDI: 4056869010106; Serial Numbers: 149100 149401 149421 149106 149202 149420 149402. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- US Nationwide.
- Lot
- UDI: 4056869010106; Serial Numbers: 149100 149401 149421 149106 149202 149420 149402
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDAArtis zee floor MN, Model Number 10094142April 27, 2022
- FDAArtis Q ceiling, Model Number 10848281April 27, 2022
Questions about this recall
Was Artis zee biplane MN, Model Number 10094143 recalled?
Yes. FDA published recall Z-0972-2022 on April 27, 2022. Reported hazard: If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
What should I do if I have this product?
Voluntary: Firm initiated · Letter