FDAFDA · Food and drugs·Recall number Z-0972-2022

Artis zee biplane MN, Model Number 10094143

Published April 27, 2022 · Recall initiated: February 23, 2022

Summary

FDA records medical device Class II recall Z-0972-2022, published April 27, 2022 and initiated February 23, 2022. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Ongoing; Reported quantity: 7 systems; Lot: UDI: 4056869010106; Serial Numbers: 149100 149401 149421 149106 149202 149420 149402. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
US Nationwide.
Lot
UDI: 4056869010106; Serial Numbers: 149100 149401 149421 149106 149202 149420 149402
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Artis zee biplane MN, Model Number 10094143 recalled?

Yes. FDA published recall Z-0972-2022 on April 27, 2022. Reported hazard: If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

What should I do if I have this product?

Voluntary: Firm initiated · Letter