Universa Firm Ureteral Stent, Global Product No. G49864 G49865 G49866 G49867 G49868 G49869 G49870 G49874 G49875 G49877 G49879 G49881 G49882 G49887 G49888
Published February 12, 2020 · Recall initiated: January 4, 2019
Summary
FDA records medical device Class II recall Z-0977-2020, published February 12, 2020 and initiated January 4, 2019. Recalling firm: Cook Inc.; Status: Terminated; Reported quantity: 13388 total; Lot: 7953153 8077981 8070702 NS8176831 7999744 7999746 8077986 8077988 8172070 8176832 8176833 8176834 NS7999747 NS8180869 7990863 8424316 8424317 8424318 NS7990867 7988039 7998403 7998404 8001365 8040131 8193749 8197930 8197931 8412989 8412990 8424332 7988039X 8003098 8506037 NS8018667 8594594 8197932 8594595 NS8197934 NS8417599 NS8417600 NS7998407 NS8033592 NS8033593. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cook Inc.
- Distributor
- Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
- Lot
- 7953153 8077981 8070702 NS8176831 7999744 7999746 8077986 8077988 8172070 8176832 8176833 8176834 NS7999747 NS8180869 7990863 8424316 8424317 8424318 NS7990867 7988039 7998403 7998404 8001365 8040131 8193749 8197930 8197931 8412989 8412990 8424332 7988039X 8003098 8506037 NS8018667 8594594 8197932 8594595 NS8197934 NS8417599 NS8417600 NS7998407 NS8033592 NS8033593
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFlexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 TheNovember 10, 2021
- FDAFlexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL0-HC, Order Number G56233 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL1-HC, Order Number G56236 Flexor Introducers andDecember 30, 2020
- FDASof-Flex Double Pigtail Ureteral Stent Set, Global Product No. G15076 G14906 G14840 G14865 G17128February 12, 2020
- FDAPacific Medical GE Corometrics Nautilus Ultrasound Transducer w/wing 5700LAX, FFCM5260February 12, 2020
Questions about this recall
Was Universa Firm Ureteral Stent, Global Product No. G49864 G49865 G49866 G49867 G49868 G49869 G49870 G49874 G49875 G49877 G49879 G49881 G49882 G49887 G49888 recalled?
Yes. FDA published recall Z-0977-2020 on February 12, 2020. Reported hazard: The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
What should I do if I have this product?
Voluntary: Firm initiated · Letter