FDAFDA · Food and drugs·Recall number Z-0978-2013

Needle Counters; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Adhesive Area (Cat. No. 3FA40SSA) Needle Counters; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Magnetic Area (Cat. No. 3FM40SSA) Product Usage: Disposal system for counting of used surgical blades and needles.

Published April 3, 2013 · Recall initiated: December 18, 2012

Summary

FDA records medical device Class II recall Z-0978-2013, published April 3, 2013 and initiated December 18, 2012. Recalling firm: Cardinal Health; Status: Terminated; Reported quantity: Cat. No. 3FA40SSA: 23,952; Cat. No. 3FM40SSA: 140,736; Lot: Catalog Number: 3FA40SSA Lot Numbers: all lot numbers less than or equal to 928782 Catalog Number: 3FM40SSA Lot Numbers: all lot numbers less than or equal to 911379. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cardinal Health
Distributor
Worldwide Distribution - USA (Nationwide) and the countried of Australia, Canada and Japan.
Lot
Catalog Number: 3FA40SSA Lot Numbers: all lot numbers less than or equal to 928782 Catalog Number: 3FM40SSA Lot Numbers: all lot numbers less than or equal to 911379
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Needle Counters; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Adhesive Area (Cat. No. 3FA40SSA) Needle Counters; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Magnetic Area (Cat. No. 3FM40SSA) Product Usage: Disposal system for counting of used surgical blades and needles. recalled?

Yes. FDA published recall Z-0978-2013 on April 3, 2013. Reported hazard: Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.

What should I do if I have this product?

Voluntary: Firm initiated · Letter