Back to search
FDAFDA · Food and drugs·Recall number Z-0978-2021

KD-620UR ELECTROSURGICAL SNARE, model no. KD-620UR - Product Usage: Used with endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract.

Published February 17, 2021 · Recall initiated: January 4, 2021

Summary

FDA records medical device Class II recall Z-0978-2021, published February 17, 2021 and initiated January 4, 2021. Recalling firm: Aomori Olympus Co., Ltd.; Status: Terminated; Reported quantity: 7,400,000 (global); 260,395 (US); Lot: 01K, 02K, 03K, 04K, 05K, 06K, 07K, 08K, 09K, 0XK, 98K, 99K, 9XK, 9YK, 9ZK. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Aomori Olympus Co., Ltd.
Distributor
US Nationwide distribution.
Lot
01K, 02K, 03K, 04K, 05K, 06K, 07K, 08K, 09K, 0XK, 98K, 99K, 9XK, 9YK, 9ZK
Risk classification
Class II
Area (FDA)
device
Country
Japan

Related recalls

Questions about this recall

Was KD-620UR ELECTROSURGICAL SNARE, model no. KD-620UR - Product Usage: Used with endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract. recalled?

Yes. FDA published recall Z-0978-2021 on February 17, 2021. Reported hazard: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.

What should I do if I have this product?

Voluntary: Firm initiated · Letter