FDAFDA · Food and drugs·Recall number Z-0982-2013

Needle Counters with Blade Disarmer; 12.2 x 10.5 x 1.7 cm; 30 Count/60 Capacity; Sterile; Foam Strip; Magnetic Area (Cat. No. 4FS60SSA) Needle Counters with Blade Disarmer; 12.2 x 10.5 x 1.7 cm; 30 Count/60 Capacity; Sterile; Double Foam Strips (Cat. No. 4FS61SSA) Needle Counters with Blade Disarmer; 12.2 x 10.5 x 1.7 cm; 30 Count/60 Capacity; Sterile; Double Numbered Magnets (Cat. No. 4MM60SSA) Product Usage: Disposal system for counting of used surgical blades and needles.

Published April 3, 2013 · Recall initiated: December 18, 2012

Summary

FDA records medical device Class II recall Z-0982-2013, published April 3, 2013 and initiated December 18, 2012. Recalling firm: Cardinal Health; Status: Terminated; Reported quantity: Cat. No. 4FS60SSA: 576; Cat. No. 4FS61SSA: 14,760; Cat. No. 4MM60SSA: 5,184; Lot: Catalog Number: 4FS60SSA Lot Numbers: all lot numbers less than or equal to 752068 Catalog Number: 4FS61SSA Lot Numbers: all lot numbers less than or equal to 802581 Catalog Number: 4MM60SSA Lot Numbers: all lot numbers less than or equal to 898715. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cardinal Health
Distributor
Worldwide Distribution - USA (Nationwide) and the countried of Australia, Canada and Japan.
Lot
Catalog Number: 4FS60SSA Lot Numbers: all lot numbers less than or equal to 752068 Catalog Number: 4FS61SSA Lot Numbers: all lot numbers less than or equal to 802581 Catalog Number: 4MM60SSA Lot Numbers: all lot numbers less than or equal to 898715
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Needle Counters with Blade Disarmer; 12.2 x 10.5 x 1.7 cm; 30 Count/60 Capacity; Sterile; Foam Strip; Magnetic Area (Cat. No. 4FS60SSA) Needle Counters with Blade Disarmer; 12.2 x 10.5 x 1.7 cm; 30 Count/60 Capacity; Sterile; Double Foam Strips (Cat. No. 4FS61SSA) Needle Counters with Blade Disarmer; 12.2 x 10.5 x 1.7 cm; 30 Count/60 Capacity; Sterile; Double Numbered Magnets (Cat. No. 4MM60SSA) Product Usage: Disposal system for counting of used surgical blades and needles. recalled?

Yes. FDA published recall Z-0982-2013 on April 3, 2013. Reported hazard: Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.

What should I do if I have this product?

Voluntary: Firm initiated · Letter