Genius 2 Tympanic Thermometer OEM Tympanic Item Code:303062
Published February 12, 2020 · Recall initiated: October 28, 2019
Summary
FDA records medical device Class II recall Z-0992-2020, published February 12, 2020 and initiated October 28, 2019. Recalling firm: Cardinal Health 200, LLC; Status: Terminated; Reported quantity: 3619 OUS; Lot: All product manufactured after October 1, 2016; Serial Numbers greater than or equal 16597907. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2®C for Genius 2 and ¿ 0.3®C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2®C for Genius 2 or +0.3®C for Genius 3.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cardinal Health 200, LLC
- Distributor
- Worldwide Distribution - US Nationwide and Foreign in the countries of: Australia Austria Belgium Canada Canary Islands Chile Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana French Polynesia Gabon Germany Gibraltar Greece Guadeloupe Hong Kong Iran, Islamic Republic Of Ireland Israel Italy Kuwait Liechtenstein Luxembourg Martinique Mayotte Mexico Netherlands New Caledonia New Zealand Norway Panama Philippines Poland Portugal Puerto Rico Qatar Reunion Romania San Marino Saudi Arabia Singapore Slovakia Slovenia South Africa South Korea Spain Swaziland Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom
- Lot
- All product manufactured after October 1, 2016; Serial Numbers greater than or equal 16597907
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump ModuleNovember 11, 2026
- FDACardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump ModuleNovember 11, 2026
- FDACardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump ModuleNovember 11, 2026
- FDACardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump ModuleNovember 11, 2026
- FDAPacific Medical GE Corometrics Nautilus Ultrasound Transducer w/wing 5700LAX, FFCM5260February 12, 2020
- FDAAccu-Chek Aviva Plus Health Network Strips- intended for Blood Glucose Monitoring System Catalog Number:February 12, 2020
Questions about this recall
Was Genius 2 Tympanic Thermometer OEM Tympanic Item Code:303062 recalled?
Yes. FDA published recall Z-0992-2020 on February 12, 2020. Reported hazard: The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2®C for Genius 2 and ¿ 0.3®C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2®C for Genius 2 or +0.3®C for Genius 3.
What should I do if I have this product?
Voluntary: Firm initiated · Letter