Back to search
FDAFDA · Food and drugs·Recall number Z-0992-2020

Genius 2 Tympanic Thermometer OEM Tympanic Item Code:303062

Published February 12, 2020 · Recall initiated: October 28, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Cardinal Health 200, LLC
Distributor
Worldwide Distribution - US Nationwide and Foreign in the countries of: Australia Austria Belgium Canada Canary Islands Chile Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana French Polynesia Gabon Germany Gibraltar Greece Guadeloupe Hong Kong Iran, Islamic Republic Of Ireland Israel Italy Kuwait Liechtenstein Luxembourg Martinique Mayotte Mexico Netherlands New Caledonia New Zealand Norway Panama Philippines Poland Portugal Puerto Rico Qatar Reunion Romania San Marino Saudi Arabia Singapore Slovakia Slovenia South Africa South Korea Spain Swaziland Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom
Lot
All product manufactured after October 1, 2016; Serial Numbers greater than or equal 16597907
Risk classification
Class II
Area (FDA)
device
Country
United States