SpaceOAR System, UPN SO-1010, SO-2101, SO-3101, and SO-4101. Used to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer to reduce the radiation dose delivered to the anterior rectum.
Published May 4, 2022 · Recall initiated: February 24, 2022
Summary
FDA records medical device Class II recall Z-1002-2022, published May 4, 2022 and initiated February 24, 2022. Recalling firm: Boston Scientific Corporation; Status: Ongoing; Reported quantity: 61,852 units; Lot: UPN SO-1010, GTIN 00864661000157, Batch Numbers: 25512955, 25512956, 25608548, 25646105, 25652246, 25676663, 25676664, 25681783, 25814685, 25814686, 25814687, 25822800, 25822812, 25822814, 25822815, 25822818, 25846481, 26059500, 26199152, 26199153, 26199154, 26199155, 26223961, 26223962, 26223963, 26223964, 26223965, 26249663, 26296567, 26296568, 26330692, 26330693, 26330694, 26330695, 26360535, 26360536, 26392679, 26450673, 26450674, 26450675, 26479093, 26514195, 26514196, 26514197, 26514198, 26552142, 26552143, 26552144, 26552145, 26552146, 26552147, 26552148, 26552149, 26552150, 26559675, 26582245, 26582246, 26582247, 26582248, 26604083, 26604084, 26604085, 26620165, 26620166, 26747082, 26747083, 26747084, 26810389, 26810390, 26840623, 26840624, 26891606, 26891607, 26942179, 26943300, 26962856, 26962857, 27092550, 27092551, 27092552, 27123363, 27164903, 27164904, 27208217, 27224418, 27251110, 27251111, 27290123, 27290124, 27290125, 27290126, 27705892, 27705893, 27777685, 27777686, 27818531, 27818831, 27860308, 28189378, 28221394, 28221395, 28221396, 28221397, 28398571, 28398572, 28413030, 28535614; UPN SO-2101, GTIN 00864661000102, Batch Numbers: 25447885, 25449325, 25454386, 25454387, 25461459, 25495940, 25498279, 25498740, 25498741, 25498742, 25503112, 25503113, 25503114, 25503115, 25503116, 25506791, 25513909, 25513910, 25513911, 25513912, 25516002, 25516004, 25518766, 25518767, 25518769, 25518770, 25604214, 25604215, 25608549, 25630371, 25630372, 25630373, 25633109, 25638880, 25638881, 25638882, 25638883, 25640316, 25640317, 25646106, 25650375, 25650376, 25652248, 25652249, 25652250, 25660251, 25660252, 25661506, 25665261, 25665262, 25665263, 25681784, 25686490, 25686491, 25692043, 25698959, 25699200, 25699201, 25699202, 25706244, 25706245, 25707436, 25707437, 25712686, 25712687, 25712688, 25712689, 25720694, 25720695, 25720696, 25725090, 25743052, 25743053, 25743054, 25746319, 25746420, 25746421, 25746422, 25746423, 25748984, 25748985, 25748986, 25748987, 25752521, 25752522, 25752523, 25752524, 25773305, 25773306, 25775424, 25775425, 25780322, 25780323, 25829631, 25829632, 25829633, 25829634, 25829635, 25838013, 25838014, 25846482, 25846483, 25846484, 25846485, 25852659, 25852880, 25852881, 25852882, 25852883, 25861746, 25861747, 25876396, 25876397, 25876398, 25876399, 25876840, 25876841, 25878667, 25878668, 25878669, 25883155, 25883156, 25883157, 25883158, 25883159, 25889579, 25890040, 25890041, 25890042, 25891035, 25907181, 25907182, 25909720, 25909721, 25909722, 25909723, 25916278, 25916279, 25916440, 25916441, 25916442, 25916443, 25921966, 25921967, 25921968, 25921969, 25921970, 25934785, 25934786, 25934787, 25934788, 25934789, 25945186, 25945187, 25945188, 25945189, 25952097, 25952098, 25952099, 25963297, 25963298, 25963299, 25963440, 25970532, 25970533, 25988372, 25988373, 25988374, 25988375, 25997188, 26005315, 26005316, 26005317, 26005318, 26014544, 26014545, 26014546, 26028461, 26028462, 26028463, 26028464, 26028465, 26035128, 26035129, 26042542, 26042543, 26042544, 26042545, 26042546, 26042547, 26042548, 26058237, 26058238, 26058239, 26059080, 26059081, 26059501, 26060392, 26071663, 26071664, 26071671, 26071672, 26071673, 26071674, 26075962, 26075997, 26078689, 26078690, 26078691, 26078692, 26078693, 26088832, 26088833, 26088834, 26088835, 26088836, 26106496, 26106497, 26106498, 26106499, 26107243, 26107244, 26107245, 26107246, 26107247, 26111500, 26111501, 26111502, 26111503, 26111504, 26111505, 26121817, 26121818, 26121819, 26122100, 26128620, 26128621, 26128622, 26128623, 26137216, 26137217, 26170536, 26170537, 26172513, 26172514, 26172515, 26172516, 26172517, 26177019, 26177460, 26177461, 26178897, 26178898, 26184397, 26184398, 26184399, 26184580, 26184581, 26186561, 26186562, 26186563, 26186564, 26186565, 26187164, 26187165, 26191446, 26191447, 26191448, 26191449, 26191450, 26193736, 26193737, 26193738, 26209535, 26209536, 26209537, 26209538, 26212441, 26212442, 26212443, 26212444, 26216386, 26216387,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Boston Scientific Corporation
- Distributor
- Worldwide - US Nationwide distribution.
- Lot
- UPN SO-1010, GTIN 00864661000157, Batch Numbers: 25512955, 25512956, 25608548, 25646105, 25652246, 25676663, 25676664, 25681783, 25814685, 25814686, 25814687, 25822800, 25822812, 25822814, 25822815, 25822818, 25846481, 26059500, 26199152, 26199153, 26199154, 26199155, 26223961, 26223962, 26223963, 26223964, 26223965, 26249663, 26296567, 26296568, 26330692, 26330693, 26330694, 26330695, 26360535, 26360536, 26392679, 26450673, 26450674, 26450675, 26479093, 26514195, 26514196, 26514197, 26514198, 26552142, 26552143, 26552144, 26552145, 26552146, 26552147, 26552148, 26552149, 26552150, 26559675, 26582245, 26582246, 26582247, 26582248, 26604083, 26604084, 26604085, 26620165, 26620166, 26747082, 26747083, 26747084, 26810389, 26810390, 26840623, 26840624, 26891606, 26891607, 26942179, 26943300, 26962856, 26962857, 27092550, 27092551, 27092552, 27123363, 27164903, 27164904, 27208217, 27224418, 27251110, 27251111, 27290123, 27290124, 27290125, 27290126, 27705892, 27705893, 27777685, 27777686, 27818531, 27818831, 27860308, 28189378, 28221394, 28221395, 28221396, 28221397, 28398571, 28398572, 28413030, 28535614; UPN SO-2101, GTIN 00864661000102, Batch Numbers: 25447885, 25449325, 25454386, 25454387, 25461459, 25495940, 25498279, 25498740, 25498741, 25498742, 25503112, 25503113, 25503114, 25503115, 25503116, 25506791, 25513909, 25513910, 25513911, 25513912, 25516002, 25516004, 25518766, 25518767, 25518769, 25518770, 25604214, 25604215, 25608549, 25630371, 25630372, 25630373, 25633109, 25638880, 25638881, 25638882, 25638883, 25640316, 25640317, 25646106, 25650375, 25650376, 25652248, 25652249, 25652250, 25660251, 25660252, 25661506, 25665261, 25665262, 25665263, 25681784, 25686490, 25686491, 25692043, 25698959, 25699200, 25699201, 25699202, 25706244, 25706245, 25707436, 25707437, 25712686, 25712687, 25712688, 25712689, 25720694, 25720695, 25720696, 25725090, 25743052, 25743053, 25743054, 25746319, 25746420, 25746421, 25746422, 25746423, 25748984, 25748985, 25748986, 25748987, 25752521, 25752522, 25752523, 25752524, 25773305, 25773306, 25775424, 25775425, 25780322, 25780323, 25829631, 25829632, 25829633, 25829634, 25829635, 25838013, 25838014, 25846482, 25846483, 25846484, 25846485, 25852659, 25852880, 25852881, 25852882, 25852883, 25861746, 25861747, 25876396, 25876397, 25876398, 25876399, 25876840, 25876841, 25878667, 25878668, 25878669, 25883155, 25883156, 25883157, 25883158, 25883159, 25889579, 25890040, 25890041, 25890042, 25891035, 25907181, 25907182, 25909720, 25909721, 25909722, 25909723, 25916278, 25916279, 25916440, 25916441, 25916442, 25916443, 25921966, 25921967, 25921968, 25921969, 25921970, 25934785, 25934786, 25934787, 25934788, 25934789, 25945186, 25945187, 25945188, 25945189, 25952097, 25952098, 25952099, 25963297, 25963298, 25963299, 25963440, 25970532, 25970533, 25988372, 25988373, 25988374, 25988375, 25997188, 26005315, 26005316, 26005317, 26005318, 26014544, 26014545, 26014546, 26028461, 26028462, 26028463, 26028464, 26028465, 26035128, 26035129, 26042542, 26042543, 26042544, 26042545, 26042546, 26042547, 26042548, 26058237, 26058238, 26058239, 26059080, 26059081, 26059501, 26060392, 26071663, 26071664, 26071671, 26071672, 26071673, 26071674, 26075962, 26075997, 26078689, 26078690, 26078691, 26078692, 26078693, 26088832, 26088833, 26088834, 26088835, 26088836, 26106496, 26106497, 26106498, 26106499, 26107243, 26107244, 26107245, 26107246, 26107247, 26111500, 26111501, 26111502, 26111503, 26111504, 26111505, 26121817, 26121818, 26121819, 26122100, 26128620, 26128621, 26128622, 26128623, 26137216, 26137217, 26170536, 26170537, 26172513, 26172514, 26172515, 26172516, 26172517, 26177019, 26177460, 26177461, 26178897, 26178898, 26184397, 26184398, 26184399, 26184580, 26184581, 26186561, 26186562, 26186563, 26186564, 26186565, 26187164, 26187165, 26191446, 26191447, 26191448, 26191449, 26191450, 26193736, 26193737, 26193738, 26209535, 26209536, 26209537, 26209538, 26212441, 26212442, 26212443, 26212444, 26216386, 26216387,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;October 21, 2026
- FDAImager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/038 BX5;September 30, 2026
- FDABoston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVISeptember 23, 2026
- FDALUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model NumberSeptember 9, 2026
- FDAda Vinici X da Vinci Xi Sureform 45 REF 480445 da Vinci X da Vinci Xi Sureform 45 Curved-Tip REF 480545 daMay 4, 2022
- FDASpaceOAR Vue System, UPN SV-1010 and SV-2101. Used to temporarily position the anterior rectal wall away fromMay 4, 2022
Questions about this recall
Was SpaceOAR System, UPN SO-1010, SO-2101, SO-3101, and SO-4101. Used to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer to reduce the radiation dose delivered to the anterior rectum. recalled?
Yes. FDA published recall Z-1002-2022 on May 4, 2022. Reported hazard: Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.
What should I do if I have this product?
Voluntary: Firm initiated · Letter