Smiths Medical, BCI® Remote Alarm Cable (Normally Open & Normally Closed), Reorder Numbers: WW3398NO & WW3398NC. For use with respiration carbon dioxide measurement equipment. Product Usage: The BCI® Remote Alarm Cable is designed for use with the 9004 Capnocheck® Capnography (Plus or Sleep) system to transfer alarm signals from the Capnocheck® Capnography system to a remote alarm system. The BCI® Remote Alarm Cable and the 9004 Capnocheck® Capnography systems are intended for use in all critical environments including ventilatory applications, patient transport, and anesthesia.
Published April 10, 2013 · Recall initiated: February 6, 2013
Summary
FDA records medical device Class I recall Z-1014-2013, published April 10, 2013 and initiated February 6, 2013. Recalling firm: Smiths Medical ASD, Inc.; Status: Terminated; Reported quantity: 39; Lot: Lot Numbers: WW3398NO0820, WW3398NO0847, WW3398NO0948, WW3398NC0822, WW3398NC0847, WW3398NC0947. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
In rare instances, the BCI® Remote Alarm Cables (BCI® Cable) are not transferring alarms when used with some remote nurse alarm systems.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Smiths Medical ASD, Inc.
- Distributor
- USA Nationwide Distribution - including the states of: CO, IL, IN, MA, SC and WI.
- Lot
- Lot Numbers: WW3398NO0820, WW3398NO0847, WW3398NO0948, WW3398NC0822, WW3398NC0847, WW3398NC0947.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACADD-Solis HPCA Ambulatory Infusion Pump; Model/Number: 21-2111-0402-09-KYO 21-2111-0402-09-KYO CADD SOLIS PIBMay 14, 2025
- FDACADD Solis VIP Ambulatory Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, MANUALMay 14, 2025
- FDAsmiths medical ProPort Low Profile Plastic Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm (5.8Fr)March 26, 2025
- FDAsmiths medical ProPort Plastic Venous Access System, Pre-assembled with PolyFlow Polyurethane Catheter, 2.6mmMarch 26, 2025
- FDAAnimas(R) 2020 Insulin Infusion Pump Product Usage: This product is indicated for continuous subcutaneousApril 10, 2013
- FDAZ-800 Volumetric Infusion System. For intravenous infusion of parenteral fluids, blood and blood products to aApril 10, 2013
Questions about this recall
Was Smiths Medical, BCI® Remote Alarm Cable (Normally Open & Normally Closed), Reorder Numbers: WW3398NO & WW3398NC. For use with respiration carbon dioxide measurement equipment. Product Usage: The BCI® Remote Alarm Cable is designed for use with the 9004 Capnocheck® Capnography (Plus or Sleep) system to transfer alarm signals from the Capnocheck® Capnography system to a remote alarm system. The BCI® Remote Alarm Cable and the 9004 Capnocheck® Capnography systems are intended for use in all critical environments including ventilatory applications, patient transport, and anesthesia. recalled?
Yes. FDA published recall Z-1014-2013 on April 10, 2013. Reported hazard: In rare instances, the BCI® Remote Alarm Cables (BCI® Cable) are not transferring alarms when used with some remote nurse alarm systems.
What should I do if I have this product?
Voluntary: Firm initiated · Letter