FDAFDA · Food and drugs·Recall number Z-1019-2013
Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewire, Models ARCSJ200W, ARCSJ260W, and ARCDJ260W. Sterile using ethylene oxide. Product Usage: The Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewires are intended to facilitate catheter placement and exchange during diagnostic or interventional procedures in the aorta, where increased support, distal flexibility, and low surface friction of the guidewire is needed.
Published April 3, 2013 · Recall initiated: March 13, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Medtronic has identified an issue involving specific lot numbers of the Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewire, Models ARCSJ200W, ARCSJ260W, and ARCDJ260W, where a change in the manufacturing process has been identified as the cause for potential kinking and breaking at the proximal weld end of the outer spring coil during in-vivo use.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Medtronic Inc. Cardiac Rhythm Disease Management
- Distributor
- Worldwide Distribution - USA (nationwide) including Puerto Rico and the countries of: Argentina, Armenia, Australia, Austria, Belgium, Bolivia, Canada, Chile, China, Colombia, Croatia, Czech Republic, Ecuador, El Salvador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Israel, Italy, Jordan, Korea, Kuwait, Latvia, Lebanon, Libya, Malaysia, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Philippines, Poland, Portugal, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Venezuela, Viet Nam.
- Lot
- Model ARCSJ200W Lot Numbers: GFWC1124, GFWC1506, GFWC2138, GFWC2652, GFWC2653, GFWD0353, GFWD0881, GFWD0882, GFWD2048, GFWD2049, GFWD2050, GFWD2051, GFWD2052, GFWE0743, GFWE0744, GFWE0745, GFWE0746, GFWE0747, GFWE3710, GFWE4272, GFWE4273, GFWE4274, GFWE4275, GFWF0143, GFWF0144, GFWF0147, GFWF0148, GFWF0149, GFWG0428, GFWG0429, GFWG0430, GFWG0431, GFWG0432, GFWG0433, GFWG0434, GFWH0333, GFWH0334, GFWH0335, GFWH0336, GFWH0337, GFWH0338, GFWI0190, GFWI0191, GFWI0192, GFWI0193, GFWI0194, GFWI0195, GFWJ0027, GFWJ0028, GFWJ0029, GFWJ2454, GFWJ2455, GFWJ2456, GFWJ2457, GFWJ2458, GFWK1910, GFWK1911, GFWL0370, GFWL2571, GFWL2572, GFXA0224. Model ARCSJ260W Lot Numbers: GFWC1501, GFWC1502, GFWC1503, GFWD0883, GFWD0884, GFWD3256, GFWE0752, GFWE0753, GFWE3711, GFWF0145, GFWF0150, GFWG0425, GFWG0426, GFWG0427, GFWH0344, GFWH0345, GFWI0201, GFWI0202, GFWI0203, GFWJ0418, GFWJ0419, GFWJ0420, GFWJ2461, GFWJ2462, GFWJ2464, GFWJ2465, GFWL0368, GFWL0369, GFWL0415, GFWL0416, GFXA0219, GFXA0220, GFXA0221, GFXA0222. Model ARCDJ260W Lot Numbers: GFWC1507, GFWC1508, GFWD0885, GFWE0750, GFWE0751, GFWF0142, GFWF0146, GFWG0423, GFWG0424, GFWH0331, GFWH0332, GFWI0186, GFWI0187, GFWI0188, GFWJ2449, GFWJ2450, GFWJ2451, GFWJ2452, GFWK1844.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States