FDAFDA · Food and drugs·Recall number Z-1024-2024
BD BBL Sensi Disc Ceftriaxone- 30 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231634 and 231635¿
Published February 14, 2024 · Recall initiated: January 8, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Becton Dickinson & Co.
- Distributor
- Domestic distribution nationwide. International distribution worldwide.
- Lot
- Catalog No. 231634 UDI-DI 00382902316343 Lots 1033870 1244758 2026381 2056573 2277488; Catalog No. 231635¿ UDI-DI 30382902316351 Lots 3062288 3158099 3254299 0344489 1033870 1088039 1116107 1148606 1180379 1211608 1244758 1335697 2026381 2028922 2056573 2180451 2277488 2277490
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States