Back to search
FDAFDA · Food and drugs·Recall number Z-1032-2024

BD BBL Sensi Disc Levofloxacin- 5 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231705 and 231706¿

Published February 14, 2024 · Recall initiated: January 8, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Becton Dickinson & Co.
Distributor
Domestic distribution nationwide. International distribution worldwide.
Lot
Catalog No. 231705 UDI-DI 00382902317050 Lots 1334510 2004644 2091272 2277179 3030830 3094117 3248344; Catalog No. 231706¿ UDI-DI 30382902317068 Lots 2277179 3030830 3094117 3184118 3248344 1334510 2004644 2091272 2213599
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.