Back to search
FDAFDA · Food and drugs·Recall number Z-1039-2025

NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10

Published February 5, 2025 · Recall initiated: January 6, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide - US Nationwide distribution in the states of CA, FL, GA, KS, MO, NE, NJ, PA, TN and the countries of CHINA, INDIA, KOREA, MALAYSIA, NETHERLANDS, SINGAPORE, TAIWAN.
Lot
UDI-DI: 00889024635746; Lot Numbers:66782843 66782840 66873137
Risk classification
Class II
Area (FDA)
device
Country
United States