FDAFDA · Food and drugs·Recall number Z-1041-2014

2.7 X 6 CORT SCREW/SELFTA 2.7 X 7 CORT SCREW/SELFTA 2.7 X 8 CORT SCREW/SELFTA 2.7 X 9 CORT SCREW/SELFTA 2.7 X 10 CORT SCREW/SELFT 2.7 X 12 CORT SCREW/SELFT 2.7 X 14 CORT SCREW/SELFT 2.7 X 16 CORT SCREW/SELFT 2.7 X 18 CORT SCREW/SELFT 2.7 X 20 CORT SCREW/SELFT 2.7 X 22 CORT SCREW/SELFT 2.7 X 24 CORT SCREW/SELFT 2.7 X 26 CORT SCREW/SELFT 2.7 X 28 CORT SCREW/SELFT 2.7 X 30 CORT SCREW/SELFT 2.7 X 32 CORT SCREW/SELFT 2.7 X 34 CORT SCREW/SELFT 2.7 X 36 CORT SCREW/SELFT 2.7 X 38 CORT SCREW/SELFT 2.7 X 40 CORT SCREW/SELFT 2.7 X 42 CORT SCREW/SELFT 2.7 X 44 CORT SCREW/SELFT 2.7 X 45 CORT SCREW/SELFT 2.7 X 46 CORT SCREW/SELFT 2.7 X 48 CORT SCREW/SELFT 2.7 X 50 CORT SCREW/SELFT 4.0 X 20 CANC SCREW/SELFT

Published February 26, 2014 · Recall initiated: October 11, 2013

Summary

FDA records medical device Class II recall Z-1041-2014, published February 26, 2014 and initiated October 11, 2013. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 7,044,680 total; Lot: 61173335 61173337 61173339 61173340 61173341 61115956 61173344 61173346 61173347 61173349 61173350 61173352 61173353 61173354 61173355 61173356 61173357 61173358 61173359 61173360 61173361 61173362 61173363 61173364 61173365 61173366 61359816. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Lot
61173335 61173337 61173339 61173340 61173341 61115956 61173344 61173346 61173347 61173349 61173350 61173352 61173353 61173354 61173355 61173356 61173357 61173358 61173359 61173360 61173361 61173362 61173363 61173364 61173365 61173366 61359816
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was 2.7 X 6 CORT SCREW/SELFTA 2.7 X 7 CORT SCREW/SELFTA 2.7 X 8 CORT SCREW/SELFTA 2.7 X 9 CORT SCREW/SELFTA 2.7 X 10 CORT SCREW/SELFT 2.7 X 12 CORT SCREW/SELFT 2.7 X 14 CORT SCREW/SELFT 2.7 X 16 CORT SCREW/SELFT 2.7 X 18 CORT SCREW/SELFT 2.7 X 20 CORT SCREW/SELFT 2.7 X 22 CORT SCREW/SELFT 2.7 X 24 CORT SCREW/SELFT 2.7 X 26 CORT SCREW/SELFT 2.7 X 28 CORT SCREW/SELFT 2.7 X 30 CORT SCREW/SELFT 2.7 X 32 CORT SCREW/SELFT 2.7 X 34 CORT SCREW/SELFT 2.7 X 36 CORT SCREW/SELFT 2.7 X 38 CORT SCREW/SELFT 2.7 X 40 CORT SCREW/SELFT 2.7 X 42 CORT SCREW/SELFT 2.7 X 44 CORT SCREW/SELFT 2.7 X 45 CORT SCREW/SELFT 2.7 X 46 CORT SCREW/SELFT 2.7 X 48 CORT SCREW/SELFT 2.7 X 50 CORT SCREW/SELFT 4.0 X 20 CANC SCREW/SELFT recalled?

Yes. FDA published recall Z-1041-2014 on February 26, 2014. Reported hazard: Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

What should I do if I have this product?

Voluntary: Firm initiated · Letter