Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.
Published January 25, 2017 · Recall initiated: July 5, 2016
Summary
FDA records medical device Class II recall Z-1043-2017, published January 25, 2017 and initiated July 5, 2016. Recalling firm: The Binding Site Group, Ltd.; Status: Terminated; Reported quantity: 809 kits; Lot: Lot #405566, 405567. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- The Binding Site Group, Ltd.
- Distributor
- U.S. distribution to the following; AR, PA, GA, NC, KY, VA, CA. Foreign distribution to the following; Uk, Germany, France, Italy, Spain, Czech Republic, Portugal, Belgium.
- Lot
- Lot #405566, 405567
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDABrand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20June 17, 2026
- FDAThe Optilite Clinical Chemistry Analyzer. IVD. Model Number: IE700. UDI-05051700017176 - Product Usage: TheMay 27, 2020
- FDAOPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565February 26, 2020
- FDABinding Site Human IgM Kit (for SPAPLUS), REF: NK012.S The device consists of reaction buffer, sheep anti-IgMDecember 19, 2018
- FDASmith & Nephew, RENASYS GO Negative Pressure Wound Therapy External Power Supply Unit, Rx only. Intended forJanuary 25, 2017
- FDAAmbra PACS UDI: +AMBRHEALTHSOLUTIONS0/$$+7_V3.16.13.0R Software Version Number: V3.16.13.0 Intended for use asJanuary 25, 2017
Questions about this recall
Was Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit. recalled?
Yes. FDA published recall Z-1043-2017 on January 25, 2017. Reported hazard: Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards.
What should I do if I have this product?
Voluntary: Firm initiated · Letter