FDAFDA · Food and drugs·Recall number Z-1045-2013
Zimmer NexPosure, retractor portals of the NexPosure MIS Access System, Part Numbers or REF numbers: N1830R80-100, N1830R80-090¿, N1830R80-080¿, N1830R80-070¿, N1830R80-060, N1830R50-100, N1830R50-090, N1830R50-080, ¿ N1830R50-070, N1830R50-060, N1830L80-100, N1830L80-090, N1830L80-080, N1830L80-070, N1830L80-060, N1830L50-100, N1830L50-090, N1830L50-080, N1830L50-070, N1830L50-060, ¿ N182450-085, N182450-075, N182450-065, N182450-055, N182450-045.¿ Retractor access system that allows placement of posterior spinal fixation implants.
Published April 10, 2013 · Recall initiated: February 8, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Possibility that the inner pouch may be compromised. The product is packaged in two pouches - a sterile inner pouch within an external pouch. NexPosure Retractors contained in a breached inner pouch have an increased risk of contamination (i.e.; loss of product sterility) if proper aseptic technique is not followed when transferring the inner package into the sterile field. Patient infection may o
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Nationwide Distribution including the states of AZ, CO, DE, FL, ID, LMA, MI, NH, NJ, NM, PA, RI, TX, and VA.
- Lot
- Lot Number 61225572 61664689 61664689 61657007 61657015 61657015 61267745 61267745 61267750 61267750 61267753 61267753 61657016 61657016 61801814 61895685 61657017 61657018 61657018 61267756 61267756 61267759 61267759 61267760 61267760 61267761 61267761 61267761 61873348 61657019 61657019 61657031 61657031
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.