FDAFDA · Food and drugs·Recall number Z-1055-2017

Digital RID Plate Reader and Software Product Code: AD400

Published January 25, 2017 · Recall initiated: October 9, 2012

Summary

FDA records medical device Class II recall Z-1055-2017, published January 25, 2017 and initiated October 9, 2012. Recalling firm: The Binding Site Group, Ltd.; Status: Terminated; Lot: UDI 05051700000376. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

If a control ring is marked after reading, the software will not flag results that are out of the specified QC range.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
The Binding Site Group, Ltd.
Distributor
U.S. distribution to the following; MA, GA, NY, IA, CA, IL, UT, NJ, NC, MD, MN. Foreign distribution to the following; UK, Germany, Spain, France.
Lot
UDI 05051700000376
Risk classification
Class II
Area (FDA)
device
Country
United Kingdom