Chord-X Chordal Sizer is designed to assist in determining the proper length of chordal replacement prostheses to repair the mitral valve and help the surgeon select a preconfigured prosthesis, or configure the correct size using straight suture material, during mitral valve repair.
Published February 1, 2017 · Recall initiated: December 13, 2016
Summary
FDA records medical device Class II recall Z-1057-2017, published February 1, 2017 and initiated December 13, 2016. Recalling firm: On-X Life Technologies, Inc.; Status: Terminated; Reported quantity: 45 units; Lot: 72-00-47420. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Four issues were uncovered by the recalling firm's CAPA: 1) Burrs were found to be present on Left Housing, Right Housing and Barrel components, 2) Loose plastic particulate was found within sealed primary packages, 3) Loose hair was discovered within four sealed primary packages and 4) Assembly error, wherein Left and Right Housing components were acceptably press-fit together, was discovered within one sealed primary package. One of the Left Housing press-fit pins was not pressed into the associated hole of the Right Housing, resulting in a small gap in the housing assembly.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- On-X Life Technologies, Inc.
- Distributor
- CA, WA, NE, DE, OH, ME, TX
- Lot
- 72-00-47420
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOn-X Mitral Heart Valve with Conform-X Sewing Ring, Model: ONXMC-25/33May 20, 2026
- FDAAllura Xper FD20 R8.2 722028 is intended for: Cardiovascular and vascular X-ray imaging applications,February 1, 2017
- FDAOEC 9800. MDL Numbers: D222250, D141598February 1, 2017
- FDAAllura Xper FD10 722026 is intended for: Cardiovascular and vascular X-ray imaging applications, includingFebruary 1, 2017
- FDAOEC 9900 Elite. MDL Numbers: D148942, D155043February 1, 2017
- FDAPittsburgh Spice Co; Buffalo Wing Sausage, 20.25 OZ for 25 #; 1235 Clairhaven St, Pittsburgh, PA 15205February 1, 2017
Questions about this recall
Was Chord-X Chordal Sizer is designed to assist in determining the proper length of chordal replacement prostheses to repair the mitral valve and help the surgeon select a preconfigured prosthesis, or configure the correct size using straight suture material, during mitral valve repair. recalled?
Yes. FDA published recall Z-1057-2017 on February 1, 2017. Reported hazard: Four issues were uncovered by the recalling firm's CAPA: 1) Burrs were found to be present on Left Housing, Right Housing and Barrel components, 2) Loose plastic particulate was found within sealed primary packages, 3) Loose hair was discovered within four sealed primary packages and 4) Assembly error, wherein Left and Right Housing components were acceptably press-fit together, was discovered within one sealed primary package. One of the Left Housing press-fit pins was not pressed into the associated hole of the Right Housing, resulting in a small gap in the housing assembly.
What should I do if I have this product?
Voluntary: Firm initiated · Letter