Sunset Compressor Nebulizer, Model NEB100 - Product Usage: The Sunset Compressor Nebulizer is an AC-powered air nebulizer compressor system intended to provide a source of compressed air for medical purposes for use in home healthcare. It is to be used with a nebulizer kit to produce medicated aerosol particles for respiratory therapy for both children and adults.

Published April 3, 2019 · Recall initiated: August 29, 2018

Summary

FDA records medical device Class II recall Z-1058-2019, published April 3, 2019 and initiated August 29, 2018. Recalling firm: SANRAI INTERNATIONAL LLC; Status: Terminated; Reported quantity: 150 units; Lot: Lot number 180223. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The product was designed for 120V usage and is being used in South Africa with 220V. There is a potential for the device to overheat and the mouthpiece can pop off.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
SANRAI INTERNATIONAL LLC
Distributor
International distribution in the country of South Africa.
Lot
Lot number 180223
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Sunset Compressor Nebulizer, Model NEB100 - Product Usage: The Sunset Compressor Nebulizer is an AC-powered air nebulizer compressor system intended to provide a source of compressed air for medical purposes for use in home healthcare. It is to be used with a nebulizer kit to produce medicated aerosol particles for respiratory therapy for both children and adults. recalled?

Yes. FDA published recall Z-1058-2019 on April 3, 2019. Reported hazard: The product was designed for 120V usage and is being used in South Africa with 220V. There is a potential for the device to overheat and the mouthpiece can pop off.

What should I do if I have this product?

Voluntary: Firm initiated · Letter